Resistance calibration
Follow these steps for resistance calibration:
1.
Set the calibrator to the resistance mode, and turn on external sensing.
Use external sense (4-wire Ω ) when calibrating all resistance ranges. Be sure that the
NOTE
calibrator external sense mode is turned on.
2.
Perform the calibration steps summarized in Table 2-10. For each step:
•
Set the calibrator to the indicated resistance, and make sure the unit is in operate.
(Use the recommended resistance or the closest available value.)
•
Send the indicated programming command. (Change the command parameter if
you are using a different calibration resistance than that shown.)
•
Wait until the Model 2015 completes each step before continuing.
Table 2-10
Resistance calibration programming steps
Calibration
step
1kΩ
10kΩ
100kΩ
1MΩ
* Use exact calibrator resistance value for parameter.
Calibrator
resistance
Calibration command*
1kΩ
:CAL:PROT:DC:STEP6 1E3
10kΩ
:CAL:PROT:DC:STEP7 10E3
100kΩ
:CAL:PROT:DC:STEP8 100E3
1MΩ
:CAL:PROT:DC:STEP9 1E6
Calibration
2-19
Parameter range
900 to 1.1E3
9E3 to 11E3
90E3 to 110E3
900E3 to 1.1E6