Resistance calibration

Follow these steps for resistance calibration:
1.
Set the calibrator to the resistance mode, and turn on external sensing.
Use external sense (4-wire Ω ) when calibrating all resistance ranges. Be sure that the
NOTE
calibrator external sense mode is turned on.
2.
Perform the calibration steps summarized in Table 2-10. For each step:
Set the calibrator to the indicated resistance, and make sure the unit is in operate.
(Use the recommended resistance or the closest available value.)
Send the indicated programming command. (Change the command parameter if
you are using a different calibration resistance than that shown.)
Wait until the Model 2015 completes each step before continuing.
Table 2-10

Resistance calibration programming steps

Calibration
step
1kΩ
10kΩ
100kΩ
1MΩ
* Use exact calibrator resistance value for parameter.
Calibrator
resistance
Calibration command*
1kΩ
:CAL:PROT:DC:STEP6 1E3
10kΩ
:CAL:PROT:DC:STEP7 10E3
100kΩ
:CAL:PROT:DC:STEP8 100E3
1MΩ
:CAL:PROT:DC:STEP9 1E6
Calibration
2-19
Parameter range
900 to 1.1E3
9E3 to 11E3
90E3 to 110E3
900E3 to 1.1E6