Date; Ndue - Keithley 2015 Service Manual

Thd multimeter
Hide thumbs Also See for 2015:
Table of Contents

Advertisement

B-8
Calibration Reference

:DATE

(:CALibration:PROTected:DATE)
Purpose
Format
Parameters
Query format :cal:prot:date?
Response
Description
NOTE
Examples

:NDUE

:CALibration:PROTected:NDUE)
Purpose
Format
Parameters
Query format :cal:prot:ndue?
Response
Description
NOTE
Examples
To send the calibration date to the instrument.
:cal:prot:date <year>, <month>, <day>
<year> = 1998 to 2097
<month> = 1 to 12
<day> = 1 to 31
<year>, <month>, <day>
The :DATE command allows you to store the calibration date in instrument
memory for future reference. You can read back the date from the instrument
over the bus by using the :DATE? query or the CALIBRATION selection in
the front panel CAL menu.
The year, month, and day parameters must be delimited by commas.
:CAL:PROT:DATE 1998,12,16
:CAL:PROT:DATE?
To send the next calibration due date to the instrument.
:cal:prot:ndue <year>, <month>, <day>
<year> = 1998 to 2097
<month> = 1 to 12
<day> = 1 to 31
<year>, <month>, <day>
The :NDUE command allows you to store the date when calibration is next
due in instrument memory. You can read back the next due date from the
instrument over the bus by using the :NDUE? query or the front panel CAL
menu.
The next due date parameters must be delimited by commas.
:CAL:PROT:NDUE 1998,12,16
:CAL:PROT:NDUE?
Send cal date (12/16/98).
Request cal date.
Send due date (12/16/98).
Request due date.