Verifying AC current

Check AC current accuracy by applying accurate AC voltage current at specific frequencies
from the AC current calibrator to the Model 2015 input and verifying that the displayed readings
fall within specified limits. Follow these steps to verify AC current:
1.
Figure 1-4
Connections for AC current verification
Model 2015
REM
STEP
SCAN
CH1
CH2
CH3
CH4
CH5
CH6
CH7
TALK
LSTN
SRQ
SHIFT
TIMER
HOLD
TRIG
FAST
MED
SLOW
REL
FILT
AUTO
ERR
MATH
T H D
dBm
dB
CONT
Ω2
Ω4
DCV
ACV
ACI
SHIFT
DCI
DELAY
HOLD
LIMITS
ON/OFF
TEST
CAL
LOCAL
TRIG
EX TRIG
STORE
RECALL
FILTER
REL
POWER
SAVE
SETUP
RS232
CONFIG
HALT
GPIB
SOURCE MEAS
STEP
SCAN
DIGITS RATE
THD
2.
3.
4.
Table 1-5

ACI limits

ACV
range
1A
3A
Connect the Model 2015 AMPS and INPUT LO jacks to the calibrator as shown in
Figure 1-4.
SENSE
INPUT
Ω 4 WIRE
HI
CH8
CH9
CH10
MATH
REAR
4W
350V
1000V
BUFFER
STAT
!
PEAK
PEAK
2015 THD MULTIMETER
LO
500V
PERIOD TCOUPL
PEAK
INPUTS
FREQ
TEMP
RANGE
F
R
AUTO
FRONT/REAR
3A 250V
AMPS
RANGE
EXIT
ENTER
Amps
Select the AC current function by pressing the ACI key, then choose the SLOW integra-
tion rate with the RATE key.
Set the Model 2015 for the 1A range.
Source 1A and 2.2A, 1kHz full-scale AC currents as summarized in Table 1-5, and
verify that the readings are within stated limits.
Applied AC voltage
1.000000A
2.20000A
Output HI
Input
LO
Output
LO
Reading limits @ 1kHz (1 year, 18°C-28°C)
0.99860 to 1.00140A
2.1949 to 2.2051A
Performance Verification
AC Current Calibrator
1-13