AC voltage calibration

Follow these steps for AC voltage calibration:
1.
Connect the calibrator to the Model 2015 INPUT HI and LO terminals as shown in
Figure 2-4.
2.
Perform the calibration steps summarized in Table 2-12. For each step:
Set the calibrator to the indicated voltage and frequency, and make sure the unit is
in operate. (You must use the stated voltage and frequency.)
Send the indicated programming command.
Wait until the Model 2015 completes each step before continuing.
Table 2-12

AC voltage calibration programming steps

Calibration step
10mV AC at 1kHz
100mV AC at 1kHz
100mV AC at 50kHz
1VAC at 1kHz
1VAC at 50kHz
10VAC at 1kHz
10VAC at 50kHz
100VAC at 1kHz
100VAC at 50kHz
700VAC at 1kHz
Calibrator voltage,
frequency
10.00000mV, 1kHz
100.0000mV, 1kHZ
100.0000mV, 50kHz
1.000000V, 1kHz
1.000000V, 50kHz
10.00000V, 1kHz
10.00000V, 50kHz
100.0000V, 1kHz
100.0000V, 50kHz
700.000V, 1kHz
Calibration
Calibration command
:CAL:PROT:AC:STEP1
:CAL:PROT:AC:STEP2
:CAL:PROT:AC:STEP3
:CAL:PROT:AC:STEP4
:CAL:PROT:AC:STEP5
:CAL:PROT:AC:STEP6
:CAL:PROT:AC:STEP7
:CAL:PROT:AC:STEP8
:CAL:PROT:AC:STEP9
:CAL:PROT:AC:STEP10