2-20
Calibration
DC current calibration
After the 1MΩ resistance point has been calibrated, follow these steps for DC current
calibration:
1.
2.
NOTE
Table 2-11
DC current calibration programming steps
Calibration
step
10mA
100mA
1A
* Change parameter if using different current.
Connect the calibrator to the AMPS and INPUT LO terminals of the Model 2015 as
shown in Figure 2-3.
Perform the calibration steps listed in Table 2-11. For each step:
•
Set the calibrator to the indicated current, and make sure the unit is in operate. (Use
the recommended current if possible.)
•
Send the indicated programming command. (Change the current parameter if you
are using a different calibration current.)
•
Wait until the Model 2015 completes each step before continuing.
If you are performing DC-only calibration, proceed to Programming calibration
dates.
Calibrator
current
10.00000mA
100.00000mA
1.000000A
Calibration command*
:CAL:PROT:DC:STEP10 10E-3
:CAL:PROT:DC:STEP11 100E-3
:CAL:PROT:DC:STEP12 1
Parameter range
9E-3 to 11E-3
90E-3 to 110E-3
0.9 to 1.1