DC current calibration

After the 1MΩ resistance point has been calibrated, follow these steps for DC current
calibration:
1.
2.
NOTE
Table 2-11

DC current calibration programming steps

Calibration
step
10mA
100mA
1A
* Change parameter if using different current.
Connect the calibrator to the AMPS and INPUT LO terminals of the Model 2015 as
shown in Figure 2-3.
Perform the calibration steps listed in Table 2-11. For each step:
Set the calibrator to the indicated current, and make sure the unit is in operate. (Use
the recommended current if possible.)
Send the indicated programming command. (Change the current parameter if you
are using a different calibration current.)
Wait until the Model 2015 completes each step before continuing.
If you are performing DC-only calibration, proceed to Programming calibration
dates.
Calibrator
current
10.00000mA
100.00000mA
1.000000A
Calibration command*
:CAL:PROT:DC:STEP10 10E-3
:CAL:PROT:DC:STEP11 100E-3
:CAL:PROT:DC:STEP12 1
Parameter range
9E-3 to 11E-3
90E-3 to 110E-3
0.9 to 1.1