Performance Verification

Verifying resistance

Check resistance by connecting accurate resistance values to the Model 2015 and verifying
that its resistance readings are within the specified limits.
CAUTION
Follow these steps to verify resistance accuracy:
1.
Figure 1-5
Connections for resistance verification (100 Ω -10M Ω ranges)
Model 2015
REM
STEP
SCAN
CH1
CH2
CH3
CH4
CH5
CH6
CH7
TALK
LSTN
SRQ
SHIFT
TIMER
HOLD
TRIG
FAST
MED
SLOW
REL
FILT
AUTO
ERR
MATH
T H D
dBm
dB
CONT
Ω2
Ω4
SHIFT
DCV
ACV
DCI
ACI
DELAY
HOLD
LIMITS
ON/OFF
TEST
CAL
LOCAL
EX TRIG
TRIG
STORE
RECALL
FILTER
REL
POWER
SAVE
SETUP
CONFIG
HALT
GPIB
RS232
SOURCE MEAS
STEP
SCAN
DIGITS RATE
THD
2.
3.
4.
5.
Do not apply more than 1100V peak between INPUT HI and LO or more
than 350V peak between SENSE HI and LO, or instrument damage could
occur .
Using shielded 4-wire connections, connect the Model 2015 INPUT and SENSE jacks
to the calibrator as shown in Figure 1-5.
Sense HI
SENSE
INPUT
Ω 4 WIRE
HI
CH8
CH9
CH10
MATH
REAR
4W
350V
1000V
BUFFER
STAT
!
PEAK
PEAK
2015 THD MULTIMETER
LO
500V
PERIOD TCOUPL
PEAK
INPUTS
FREQ
TEMP
RANGE
F
R
AUTO
FRONT/REAR
3A 250V
AMPS
RANGE
EXIT
ENTER
Sense LO
Set the calibrator for 4-wire resistance with external sense on.
Select the Model 2015 4-wire resistance function by pressing the Ω 4 key, then choose
the SLOW integration rate with the RATE key.
Set the Model 2015 for the 100 Ω range, and make sure the FILTER is on.
Recalculate reading limits based on actual calibrator resistance values.
Sense HI
Input
HI
Output
HI
Input
Output
LO
LO
Sense LO
Note: Use shielded low-thermal cables to
minimize noise. Enable or disable
calibrator external sense as indicated
in procedure.
Resistance Calibrator

Advertisement

Table of Contents
loading

Table of Contents