Verifying total harmonic distortion

Follow the steps below to verify the Model 2015 total harmonic distortion function.
1.
Figure 1-8
Connections for total harmonic distortion verification
Model 2015
REM
STEP
SCAN
CH1
CH2
CH3
CH4
CH5
CH6
CH7
CH8
TALK
LSTN
SRQ
SHIFT
TIMER
HOLD
TRIG
FAST
MED
SLOW
REL
FILT
AUTO
ERR
MATH
T H D
dBm
dB
CONT
Ω2
Ω4
SHIFT
DCV
ACV
DCI
ACI
DELAY
HOLD
LIMITS
ON/OFF
TEST
CAL
LOCAL
EX TRIG
TRIG
STORE
RECALL
FILTER
REL
POWER
SAVE
SETUP
CONFIG
HALT
GPIB
RS232
SOURCE MEAS
STEP
SCAN
DIGITS RATE
THD
2.
3.
4.
5.
6.
Connect the low-distortion function generator to the Model 2015 INPUT jacks. (See
Figure 1-8.)
BNC-to-Dual
Banana Plug
Adapter
SENSE
INPUT
Ω 4 WIRE
HI
CH9
CH10
MATH
REAR
4W
350V
1000V
BUFFER
STAT
!
PEAK
PEAK
2015THD MULTIMETER
LO
500V
PERIOD TCOUPL
PEAK
INPUTS
FREQ
TEMP
RANGE
F
R
AUTO
FRONT/REAR
3A 250V
AMPS
RANGE
EXIT
ENTER
Set the function generator to output a 1kHz, 0.95V RMS sine wave with an unbalanced,
high-impedance output.
Using the MEAS key, set the following operating modes:
TYPE: THD
FREQ: AUTO
UPR HARM: 10
UNITS: PERC
SFIL: NONE
Select the Model 2015 THD function by pressing SHIFT then THD.
Use the down RANGE key to select the 1V range.
Verify that the Model 2015 THD reading is <0.004%.
Low Distortion Function Generator