:STEP11

(CALibration:PROTected:DC:STEP11)
Purpose
Format
Parameter
Description
Example

:STEP12

(CALibration:PROTected:DC:STEP12)
Purpose
Format
Parameter
Description
Example
To program the 100mA comprehensive calibration step.
:cal:prot:dc:step11 <Cal_current>
<Cal_current> = 90E-3 to 110E-3 [A]
:STEP11 programs the 100mA comprehensive calibration step. The allow-
able range of the calibration current parameter is from 90E-3 to 110E-3. Use
the 100E-3 value whenever possible for best results.
:CAL:PROT:DC:STEP11 0.1
To program the 1A comprehensive calibration step.
:cal:prot:dc:step12 <Cal_current>
<Cal_current> = 0.9 to 1.1 [A]
:STEP12 programs the 1A comprehensive calibration step. The allowable
range of the calibration current parameter is from 0.9 to 1.1. Use a value of
1 whenever possible for best results.
:CAL:PROT:DC:STEP12 1
Calibration Reference
Program 100mA step.
Program 1A step.