A-6
Specifications
TRUE RMS AC VOLTAGE AND CURRENT CHARACTERISTICS
ACCURACY
1
: ±(% of reading + % of range), 23°C ±5 °C
VOLTAGE
RANGE
RESOLUTION
100.0000 mV
0.1 µV
1.000000
V
1.0 µV
10.00000
V
10 µV
100.0000
V
100 µV
750.000
V
1 mV
CURRENT
RANGE
RESOLUTION
1.000000
A
1 µA
3.00000
A
10 µA
HIGH CREST FACTOR ADDITIONAL ERROR ±(% of reading)
CREST FACTOR:
1–2
ADDITIONAL ERROR:
0.05
AC OPERATING CHARACTERISTICS
FUNCTION
DIGITS
ACV (all ranges), and
6
1
2 3
ACI (all ranges)
6
1
2 3
6
1
2 4
1
2 3
6
6
1
2 4
ADDITIONAL LOW FREQUENCY ERRORS ±(% of reading)
20Hz –
30Hz
30Hz –
50Hz
50Hz – 100Hz
100Hz – 200Hz
200Hz – 300Hz
> 300Hz
AC SYSTEM SPEEDS
2, 5
FUNCTION/RANGE CHANGE
6
AUTORANGE TIME: <3 s.
ASCII READINGS TO RS-232 (19.2K BAUD)
MAX. INTERNAL TRIGGER RATE
MAX. EXTERNAL TRIGGER RATE
CALIBRATION
CYCLE
90 Days
1 Year
TEMPERATURE
COEFFICIENT
8
CALIBRATION
CYCLE
90 Day/1 Year
90 Day/1 Year
TEMPERATURE
8
COEFFICIENT
2–3
3–4
0.15
0.30
2
READINGS/s
RATE
2s/reading
SLOW
1.4
MED
4.8
MED
2.2
FAST
35
FAST
SLOW
MED
0
0.3
0
0
0
0
0
0
0
0
0
0
: 4 / s.
4
: 50 / s.
4
: 300 / s.
4
: 260 / s.
3 Hz–
10 Hz–
10 Hz
20 kHz
0.35 + 0.03
0.05 + 0.03
0.35 + 0.03
0.06 + 0.03
0.035 + 0.003
0.005 + 0.003
3 Hz -
10 Hz -
10 Hz
5 kHz
0.30 + 0.04
0.10 + 0.04
0.35 + 0.06
0.15 + 0.06
0.035 + 0.006
0.015 + 0.006
7
4–5
0.40
BANDWIDTH
3 Hz–300 kHz
30 Hz–300 kHz
30 Hz–300 kHz
300 Hz–300 kHz
300 Hz–300 kHz
FAST
1.0
0.18
0.10
0
20 kHz–
50 kHz–
50 kHz
100 kHz
0.11 + 0.05
0.60 + 0.08
0.12 + 0.05
0.60 + 0.08
0.006 + 0.005
0.01 + 0.006
100 kHz–
300 kHz
4 + 0.5
4 + 0.5
0.03 + 0.01

Advertisement