Verifying DC current

Check DC current accuracy by applying accurate DC currents from the DC current calibrator
to the AMPS input of the Model 2015 and verifying that the displayed readings fall within spec-
ified limits.
Follow these steps to verify DC current accuracy:
1.
Figure 1-3
Connections for DC current verification
Model 2015
REM
STEP
SCAN
CH1
CH2
CH3
CH4
CH5
CH6
CH7
TALK
LSTN
SRQ
SHIFT
TIMER
HOLD
TRIG
FAST
MED
SLOW
REL
FILT
AUTO
ERR
MATH
T H D
dBm
dB
CONT
Ω2
Ω4
DCV
ACV
ACI
SHIFT
DCI
DELAY
HOLD
LIMITS
ON/OFF
TEST
CAL
LOCAL
TRIG
EX TRIG
STORE
RECALL
FILTER
REL
POWER
SAVE
SETUP
RS232
CONFIG
HALT
GPIB
SOURCE MEAS
STEP
SCAN
DIGITS RATE
THD
2.
3.
4.
Table 1-4

DCI limits

DCI
range
10mA
100mA
1A
3A
*Source positive and negative currents with values shown.
Connect the Model 2015 AMPS and INPUT LO jacks to the calibrator as shown in
Figure 1-3.
SENSE
INPUT
Ω 4 WIRE
HI
CH8
CH9
CH10
MATH
REAR
4W
350V
1000V
BUFFER
STAT
!
PEAK
PEAK
2015 THD MULTIMETER
LO
500V
PERIOD TCOUPL
PEAK
INPUTS
FREQ
TEMP
RANGE
F
R
AUTO
FRONT/REAR
3A 250V
AMPS
RANGE
EXIT
ENTER
Select the DC current measurement function by pressing the DCI key, then choose the
SLOW integration rate with the RATE key.
Set the Model 2015 for the 10mA range.
Source positive and negative full-scale currents for each of the ranges listed in Table
1-4, and verify that the readings for each range are within stated limits.
Applied DC current*
10.0000mA
100.0000mA
1.000000A
2.20000A
Input
LO
Output HI
Amps
Output
Note: Be sure calibrator is set for
LO
normal current output.
Reading limits (1 year, 18°C-28°C)
9 .99460 to 10.00540mA
99.9100 to 100.0900mA
0.999160 to 1.000840A
2.197315 to 2.202685A
DC Current Calibrator