Performance Verification

Verifying DC voltage

Check DC voltage accuracy by applying accurate voltages from the DC voltage calibrator to
the Model 2015 INPUT jacks and verifying that the displayed readings fall within specified
limits.
CAUTION
Follow these steps to verify DC voltage accuracy:
1.
NOTE
Figure 1-1
Connections for DC volts verification
Model 2015
REM
STEP
SCAN
CH1
CH2
CH3
CH4
CH5
CH6
CH7
TALK
LSTN
SRQ
SHIFT
TIMER
HOLD
TRIG
FAST
MED
SLOW
REL
FILT
AUTO
ERR
dBm
MATH
T H D
dB
CONT
Ω2
Ω4
SHIFT
DCV
ACV
DCI
ACI
DELAY
HOLD
LIMITS
ON/OFF
TEST
CAL
LOCAL
EX TRIG
TRIG
STORE
RECALL
FILTER
REL
POWER
SAVE
SETUP
CONFIG
HALT
GPIB
RS232
SOURCE MEAS
STEP
SCAN
DIGITS RATE
THD
Do not exceed 1100V peak between INPUT HI and INPUT LO because
instrument damage may occur.
Connect the Model 2015 HI and LO INPUT jacks to the DC voltage calibrator as shown
in Figure 1-1.
Use shielded, low-thermal connections when testing the 100mV and 1V ranges to
avoid errors caused by noise or thermal effects. Connect the shield to the calibrator's
output LO terminal.
SENSE
Ω 4 WIRE
INPUT
HI
MATH
CH8
CH9
CH10
REAR
4W
350V
1000V
BUFFER
STAT
!
PEAK
PEAK
2015 THD MULTIMETER
LO
500V
PERIOD TCOUPL
PEAK
INPUTS
FREQ
TEMP
RANGE
F
R
AUTO
FRONT/REAR
3A 250V
AMPS
RANGE
EXIT
ENTER
Input HI
Output HI
Input
LO
Output
LO
Note: Use shielded, low-thermal cables
DC Voltage Calibrator
for 100mV and 1V ranges.