2-10
Calibration
3.
NOTE
Table 2-3
DC volts calibration summary
Calibration
step
+10V
-10V
100V
Perform the steps listed in Table 2-3 to complete DC volts calibration. For each calibra-
tion step:
•
Set the calibrator to the indicated value, and make sure it is in operate.
•
Press the ENTER key to calibrate that step.
•
Wait until the Model 2015 finishes each step. (The unit will display the CALI-
BRATING message while calibrating.)
If your calibrator cannot output the values recommended in Table 2-3, use the left and
right arrow keys, and the up and down range keys to set the Model 2015 display value
to match the calibrator output voltage.
Calibrator
voltage
+10.00000V
-10.00000V
+100.0000V
Allowable range
+9V to +11V
-9V to -11V
+90V to +110V