Dc Volts Calibration Summary - Keithley 2015 Service Manual

Thd multimeter
Hide thumbs Also See for 2015:
Table of Contents

Advertisement

2-10
Calibration
3.
NOTE
Table 2-3

DC volts calibration summary

Calibration
step
+10V
-10V
100V
Perform the steps listed in Table 2-3 to complete DC volts calibration. For each calibra-
tion step:
Set the calibrator to the indicated value, and make sure it is in operate.
Press the ENTER key to calibrate that step.
Wait until the Model 2015 finishes each step. (The unit will display the CALI-
BRATING message while calibrating.)
If your calibrator cannot output the values recommended in Table 2-3, use the left and
right arrow keys, and the up and down range keys to set the Model 2015 display value
to match the calibrator output voltage.
Calibrator
voltage
+10.00000V
-10.00000V
+100.0000V
Allowable range
+9V to +11V
-9V to -11V
+90V to +110V