B-14
Calibration Reference

AC calibration commands

The :AC commands perform comprehensive (user) calibration of the ACV and ACI functions.
Table B-3 summarizes these calibration commands.
Table B-3

AC calibration commands

Command
:CALibration
:PROTected
:AC
:STEP1
:STEP2
:STEP3
:STEP4
:STEP5
:STEP6
:STEP7
:STEP8
:STEP9
:STEP10
:STEP11
:STEP12
:STEP13
Description
10mV AC at 1kHz calibration step.
100mV AC at 1kHz calibration step.
100mV AC at 50kHz calibration step.
1V AC at 1kHz calibration step.
1V AC at 50kHz calibration step.
10V AC at 1kHz calibration step.
10V AC at 50kHz calibration step.
100V AC at 1kHz calibration step.
100V AC at 50kHz calibration step.
700V AC at 1kHz calibration step.
100mA AC at 1kHz calibration step.
1A AC at 1kHz calibration step.
2A AC at 1kHz calibration step.

Advertisement