Verifying function generator amplitude

Follow the steps below to verify Model 2015 function generator amplitude:
1.
Figure 1-9
Connections for function generator amplitude verification
REM
TALK
LSTN
SRQ
SHIFT
TIMER
MATH
SHIFT
DCV
DELAY
LOCAL
EX TRIG
POWER
SAVE
SOURCE MEAS
THD
2.
3.
4.
Connect the rear panel SOURCE OUTPUT jack to the front panel INPUT jacks. (See
Figure 1-9.)
Model 2015
STEP
SCAN
CH1
CH2
CH3
CH4
CH5
CH6
CH7
CH8
CH9
CH10
MATH
REAR
4W
HOLD
TRIG
FAST
MED
SLOW
REL
FILT
AUTO
ERR
BUFFER
STAT
2015 THD MULTIMETER
T H D
dBm
dB
CONT
PERIOD TCOUPL
Ω2
Ω4
ACV
DCI
ACI
FREQ
TEMP
RANGE
HOLD
LIMITS
ON/OFF
TEST
CAL
AUTO
TRIG
STORE
RECALL
FILTER
REL
RANGE
SETUP
CONFIG
HALT
GPIB
RS232
STEP
SCAN
DIGITS RATE
EXIT
ENTER
Use the SOURCE key to set the function generator operating modes as follows:
SINE OUT: ON
FREQ: 1kHz
IMPEDANCE: HIZ
AMPL: 4.0000V
Press the ACV key to select the AC voltage function, and choose the 10V range.
Verify that the AC voltage reading is between 3.986V and 4.014V.
Connect INPUT jacks to
rear panel SOURCE OUTPUT
SENSE
Ω 4 WIRE
INPUT
HI
350V
1000V
!
PEAK
PEAK
LO
500V
PEAK
INPUTS
F
R
FRONT/REAR
3A 250V
AMPS
Performance Verification
1-19