Table 2-8 Dc Voltage Calibration Programming Steps; Table 2-9 Resistance Calibration Programming Steps - Keithley 2701 Service Manual

Ethernet-based dmm / data acquisition system
Hide thumbs Also See for 2701:
Table of Contents

Advertisement

Model 2701 Service Manual
NOTE
Table 2-8
DC voltage calibration programming steps
Calibration step
+10V
-10V
100V
*Use recommended value where possible. Change parameter accordingly if using a different calibrator voltage.
Resistance calibration
Follow these steps for resistance calibration:
1.
NOTE
2.
Table 2-9
Resistance calibration programming steps
Calibration
step
1kΩ
10kΩ
100kΩ
1MΩ
*Use exact calibrator resistance value for parameter.
Ensure the calibrator has settled to the final value. You can do so by verifying that
the "Settled" indicator is off or by using the *OPC? (operation complete) query.
Calibrator voltage
+10.00000V
-10.00000V
100.0000V
Set the calibrator to the resistance mode and turn on external sensing.
Use external sense (4-wire Ω) when calibrating all resistance ranges. Be sure that
the calibrator external sense mode is turned on.
Perform the calibration steps summarized in
Set the calibrator to the indicated resistance and make sure the unit is in operate.
(Use the recommended resistance or the closest available value.)
Send the indicated programming command. (Change the command parameter if
you are using a different calibration resistance than that shown.)
Wait until the Model 2701 completes each step before continuing.
Calibrator
resistance
1kΩ
10kΩ
100kΩ
1MΩ
Calibration command*
:CAL:PROT:DC:STEP3 10
:CAL:PROT:DC:STEP4 -10
:CAL:PROT:DC:STEP5 100
Table
2-9. For each step:
Calibration command*
:CAL:PROT:DC:STEP6 1E3
:CAL:PROT:DC:STEP7 10E3
:CAL:PROT:DC:STEP8 100E3
:CAL:PROT:DC:STEP9 1E6
Calibration
2-15
Parameter range
9 to 11
-9 to -11
90 to 110
Parameter range
900 to 1.1E3
9E3 to 11E3
90E3 to 110E3
900E3 to 1.1E6

Advertisement

Table of Contents
loading

Table of Contents