Calibration Date - Keithley 2110 Calibration Manual

5 1/2 digit multimeter
Hide thumbs Also See for 2110:
Table of Contents

Advertisement

Section 2: Adjustment

Calibration date

1. Update the calibration date by sending the following command:
CAL:PROT:DATE <"MM/DD/YYYY">
The date <string> must be enclosed in double quotes ("<string>").
Example: If calibration date is 5 Apr 2012, send the following command:
CAL:PROT:DATE "04/05/2012"
Calibration due date is automatically calculated for 12 months.
2. Save calibration constants in EEPROM by sending the following command:
CAL:PROT:SAVE
3. Lock calibration by sending the following command:
CAL:PROT:LOCK
2-18
Model 2110 5½ Digit Multimeter Calibration Manual
2110-905-01 Rev. C / August 2013