UNITS ENABLE/DISABLE SELECTION
The units enable/disable feature controls which weighing units the scale will toggle between when the
units key is repeatedly pressed. The units of measure can be limited to two, three, four, etc. or it can
be entirely disabled so that only one unit of measure can be displayed. For example, if only the grams
and pounds units are enabled, pressing the units key repeatedly will toggle between grams and
pounds. To enable or disable units, follow this procedure:
1. Enter —12221.
2. Press the setup/cal key. The display will now read "CAL U". Press the setup/cal key again.
3. Read the display. The scale now asks if you wish to enable the 'X' units.
4. Enter 1 if you wish to enable the 'X' units (enter 0 if you wish to disable the 'X' units.)
5. Press the setup/cal key. Read the display. The scale now asks if you wish to enable grams.
6. Enter 1 if you wish to enable grams (enter 0 if you wish to disable grams.)
7. Press the setup/cal key. Read the display. The scale now asks if you wish to enable ounces.
8. Enter 1 if you wish to enable ounces (enter 0 if you wish to disable ounces.)
9. Press the setup/cal key. Read the display. The scale now asks if you wish to enable pounds.
10. Enter 1 if you wish to enable pounds (enter 0 if you wish to disable pounds.)
11. Press the setup/cal key. Read the display. The scale now asks if you wish to enable troy
ounces.
12. Enter 1 if you wish to enable troy ounces (enter 0 if you wish to disable troy ounces.)
13. Press the setup/cal key. Read the display. The scale now asks if you wish to enable penny-
weights.
14. Enter 1 if you wish to enable pennyweights (enter 0 if you wish to disable pennyweights.)
15. Press the setup/cal key. Read the display. The scale now asks if you wish to enable carats.
16. Enter 1 if you wish to enable carats (enter 0 if you wish to disable carats.)
17. Press the setup/cal key.
NOTE: If you attempt to disable all units, the scale will respond "UnAbLE". This feature affects the
units convert function by displaying only the units that are enabled in the following sequence
of units: 'X' units, grams, ounces, pounds, troy ounces, pennyweight and carats.
Pressing —, units will return the scale to the factory default units, if it is enabled, otherwise it
will display the next unit that is enabled.