Fluke CombiScope PM3370B User Manual page 96

Autoranging combiscope instrument
Hide thumbs Also See for CombiScope PM3370B:
Table of Contents

Advertisement

HOW TO USE THE INSTRUMENT
MEASURE
-
Press the 'RETURN' softkey to return to the 'MEASURE' menu.
-
Turn on MEAS 2.
MEAS 2 performs a frequency measurement on the same signal. The screen
displays this result as : 'ch1 freq= ....kHz'.
-
Remove the Probe Adjust signal from channel 1
Because of the absence of an input signal, no frequency can be measured. This
results in the display 'ch1 freq= ----Hz'.
MEASURE
-
Connect the Probe Adjust signal to channels 1 and 2.
-
Turn on channel 2.
-
Shift the signal on channel 1 to the upper half of the screen and the
signal on channel 2 to the lower half of the screen.
-
Press the softkey next to MEAS1.
The SELECT MEAS1 submenu is displayed.
-
Press the first softkey to select 'delay'
The delay measurement is
performed automatically.
Observe that the first
measurement is linked to the
channel 1 signal (on positive
slope) and the second is listed to
the channel 2 signal (on
negative slope).
The delay is displayed in the top
corner of the screen
'ch1:
del = .... s'.
-
Press the RETURN softkey to return to the MEASURE menu.
This measurement is very useful when you must compensate unequal cable delays.
1
2
DELAY
4 - 51
MEAS 2-FREQ
DELAY
SELECT
MEAS 1
volt
time
delay
-
ch1
T
ch2
ch1
ch2
-
RETURN
ST6723