Fluke P3000 Series User Manual
Fluke P3000 Series User Manual

Fluke P3000 Series User Manual

Pneumatic deadweight tester
Hide thumbs Also See for P3000 Series:
Table of Contents

Advertisement

P3000 Series
Pneumatic Deadweight Tester
Users Manual
PN 3952260
November 2010
© 2010 Fluke Corporation. All rights reserved. Printed in USA. Specifications are subject to change without notice.
Fluke-Direct
All product names are trademarks of their respective companies.
.com
info@Fluke-Direct.com
1.888.610.7664

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the P3000 Series and is the answer not in the manual?

Questions and answers

Summary of Contents for Fluke P3000 Series

  • Page 1 P3000 Series Pneumatic Deadweight Tester Users Manual PN 3952260 November 2010 © 2010 Fluke Corporation. All rights reserved. Printed in USA. Specifications are subject to change without notice. Fluke-Direct All product names are trademarks of their respective companies. .com info@Fluke-Direct.com...
  • Page 2 Fluke authorized resellers shall extend this warranty on new and unused products to end-user customers only but have no authority to extend a greater or different warranty on behalf of Fluke. Warranty support is available only if product is purchased through a Fluke authorized sales outlet or Buyer has paid the applicable international price.
  • Page 3: Table Of Contents

    Chapter Title Page 1 General Information ................1-1 Introduction......................1-1 How to Contact Fluke ..................1-1 Safety Information ..................... 1-1 Safety Summary .................... 1-1 Compressed Gas .................... 1-2 Heavy Weights ....................1-2 Personal Protective Equipment..............1-2 Symbols Used in this Manual ................1-2 Operating Principle ....................
  • Page 4 P3000 Series Users Manual Maintenance & Servicing ..............5-1 Introduction......................5-1 PCU Assembly — 10 mm Nominal Diameter — Pressure ....... 5-1 Piston Disassembly..................5-1 Piston Cleaning....................5-1 Piston Re-Assembly ..................5-2 Replacement PCU Assembly ................ 5-2 PCU Assembly — 10 mm Nominal Diameter — Vacuum ....... 5-3 Piston Disassembly..................
  • Page 5 PCU Assembly - 16 mm - Parts List ..............5-7 5-4. PCU Assembly - 22 mm - Parts List ..............5-9 5-5. Handpump Assembly - Parts List................5-11 5-6. System Selector Valve - Parts List................. 5-13 5-7. Safety Valve - Parts List ..................5-14 Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
  • Page 6 PCU Assembly - 22 mm ..................5-9 5-5. Handpump Assembly ..................... 5-11 5-6. System Selector Valve ................... 5-13 5-7. Safety Valve ......................5-14 8-1. Dirt/Moisture Trap ....................8-1 8-2. Angle Adapter ......................8-2 8-3. Pointer/Remover Punch..................8-2 Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
  • Page 7: General Information

    W Warning If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired. Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
  • Page 8: Compressed Gas

    AC (Alternating Current) Earth Ground Important Information: refer to manual Do not dispose of this product as unsorted municipal waste. Go to Fluke’s website for recycling information. Operating Principle Deadweight Testers are the primary standard for pressure measurement. Utilizing the...
  • Page 9: Operating & Storage Specification

    Some countries have geographic/geological survey/mapping organizations that have the data readily available. If not, the countries’ National Standards Laboratory may be able to recommend a source of suitable information. Temperature Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
  • Page 10: Model Variations

    TRUE PRESSURE TRUE PRESSURE Model Variations This manual covers many pressure ranges and models within the P3000 series of instruments. The following illustrations detail the physical appearance of the six basic models: Model 3015 gjn024.bmp Figure 1-1. Single PCU Instrument Fluke-Direct .com...
  • Page 11: Single Pcu Instrument With Handpump

    General Information Model Variations Model 3012P, 3013P, 3014P and 3015P gjn025.bmp Figure 1-2. Single PCU Instrument with Handpump Model 3022, 3023 and 3025 gjn026.bmp Figure 1-3. Dual PCU (Pressure & Vacuum) Instrument Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
  • Page 12: Dual Pcu (Pressure & Vacuum) Instrument With Handpump

    P3000 Series Users Manual Model 3022P, 3023P and 3025P gjn027.bmp Figure 1-4. Dual PCU (Pressure & Vacuum) Instrument with Handpump Model 3011 gjn028.bmp Figure 1-5. Vacuum Instrument Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
  • Page 13: Vacuum Instrument With Handpump

    General Information Model Variations Model 3011P gjn029.bmp Figure 1-6. Vacuum Instrument with Handpump Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
  • Page 14: Preparation

    Dirt/Moisture Trap, P5531, can be mounted to the test port of the deadweight tester. This will prevent any dirt particles or moisture droplets present within the DUT from passing into the deadweight tester system, (see Chapter 8). Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
  • Page 15: Connections

    P3000 Series Users Manual Connections Fit the device under test (DUT) to the test port using the method described below: W Warning DO NOT use Teflon/PTFE tape on these connections, as this will prevent correct sealing. The Gauge Adapter sealing system is designed for hand-tight sealing up to 20,000 psi / 1,400 bar-wrenches or similar tools are not required —...
  • Page 16: Connect Assembly To Test Port

    Connections gjn007.bmp Figure 2-2. Connect Assembly to Test Port gjn008.bmp Figure 2-3. Hand-Tighten Only 3. To adjust the position to face forward, hold the gauge adapter and turn the instrument COUNTER-CLOCKWISE, so that it faces forward. Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
  • Page 17: Adjust Gauge Position

    P3000 Series Users Manual gjn009.bmp Figure 2-4. Adjust Gauge Position 4. Hold the instrument steady, whilst turning the gauge adapter COUNTER-CLOCKWISE until it pulls down onto the O-ring. gjn010.bmp Figure 2-5. Tighten Gauge Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
  • Page 18: Test Port Insert

    Table 2-1. Test Port Inserts - Parts List Item Description Part O Ring 3865142 Test Port Insert 3919892 O Ring 3883397 Test Port 3921566 To calibrate panel-mounted gauges with pressure connections in the rear, use an Angle Adapter (see Chapter 8, Ancillary Equipment). Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
  • Page 19: Operation

    Chapter 3 Operation Introduction All P3000 Series pneumatic deadweight testers have a monitor gauge fitted to the instrument top plate, to serve as a guide to the operator by indicating the approximate pressure in the system. Single PCU Models — Pressure 1.
  • Page 20: Single Pcu Models - Vacuum

    P3000 Series Users Manual 8. Depressurize the system by SLOWLY opening the Exhaust valve, as sudden depressurization will cause the weight stack to fall quickly, which may damage the piston assembly. 9. Remove the weights from the piston. Single PCU Models — Vacuum 1.
  • Page 21: Weight Rotation

    4. Pressure Mode: Operate as per Single PCU Models - Pressure, above. 5. Vacuum Mode: Operate as per Single PCU Models - Vacuum, above. gjn013.bmp Figure 3-1. Weight Rotation Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
  • Page 22: Calibration In Different Pressure Units

    (e.g. local gravity, temperature, pressure head, etc.) to enhance the pressure measurement accuracy of the instrument. It will allow calibration in any of 12 different pressure units, using the existing weight set. Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
  • Page 23: Maintenance & Servicing

    6. After removal from the solvent, using a NEW tissue, repeat the cleaning procedure in step 4. 7. Place piston carefully on a NEW tissue where it will not be damaged while the cylinder is cleaned. Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
  • Page 24: Piston Re-Assembly

    P3000 Series Users Manual W Caution Never touch the working surface of a clean piston with bare fingers — the natural oil in your skin can cause the piston and cylinder to stick. 8. Wipe any dirt or moisture from the outside surfaces of the cylinder (2).
  • Page 25: Pcu Assembly - 10 Mm Nominal Diameter - Vacuum

    4. To remove all traces of contamination, the piston can be cleaned in a suitable solvent. W Caution O-ring seals (where fitted) are Nitrile rubber, and should not be immersed in solvents, as they will become damaged. They should be wiped carefully with a new tissue. Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
  • Page 26: Piston Re-Assembly

    P3000 Series Users Manual 5. After removal from the solvent, using a NEW tissue, repeat the cleaning procedure in step 3, taking great care to ensure that all traces of solvent are removed from the inside of the tubular piston and weight carrier.
  • Page 27: Pcu Assembly - 16 Mm Nominal Diameter

    5. To remove all traces of contamination, the piston can be cleaned in a suitable solvent. W Caution O-ring seals (where fitted) are Nitrile rubber, and should not be immersed in solvents, as they will become damaged. They should be wiped carefully with a new tissue. Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
  • Page 28: Piston Re-Assembly

    P3000 Series Users Manual 6. After removal from the solvent, using a NEW tissue, repeat the cleaning procedure in step 4, take great care to ensure that all traces of solvent are removed from inside the tubular piston. 7. Place piston carefully on a NEW tissue where it will not be damaged while the cylinder is cleaned.
  • Page 29: Pcu Assembly - 22 Mm Nominal Diameter

    5. To remove all traces of contamination, the piston can be cleaned in a suitable solvent. W Caution O-ring seals (where fitted) are Nitrile rubber, and should not be immersed in solvents, as they will become damaged. They should be wiped carefully with a new tissue. Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
  • Page 30: Piston Re-Assembly

    P3000 Series Users Manual 6. After removal from the solvent, using a NEW tissue, repeat the cleaning procedure in step 4, take great care to ensure that all traces of solvent are removed from inside the tubular piston. 7. Place piston carefully on a NEW tissue where it will not be damaged while the cylinder is cleaned.
  • Page 31: Top Plate Removal

    6. Carefully lift the top plate to reveal the inlet connection(s) from the external supply manifold (mounted on the rear panel of the case). 7. Disconnect the tubing, and lift the top plate clear of the case. Fluke-Direct .com info@Fluke-Direct.com...
  • Page 32: Handpump Assembly

    P3000 Series Users Manual W Caution When handling the top plate assembly, it is good practice to remove the piston assemblies to avoid accidental damage. Note Replacement is simply the reverse of the above procedure. Handpump Assembly Disassembly 1. Disconnect tubing from check valve (19) and/or (23).
  • Page 33: Handpump Assembly - Parts List

    3918868 O Ring 3867548 Check Valve Body Spec O Ring 3918984 Coupling 3918991 Snap Ring 3918797 Flexible Tubing 3922387 Piston Shaft Spec. Coupling 3923564 O Ring 3883480 Check Valve Assy. 3920249 Piston Head Assy. Spec. 5-11 Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
  • Page 34: System Selector Valve

    P3000 Series Users Manual System Selector Valve Before commencing disassembly of the selector valve, make a note of the tubing connections to avoid problems during reassembly. Disassembly 1. Disconnect nylon tubing from connectors (11). 2. Loosen set screws (1), and remove knob (2) from valve shaft (3).
  • Page 35: Safety Valve

    3. Withdraw safety valve (1) from safety plug (3), and inspect O ring (2) for cleanliness or signs of damage—replace as necessary. Re-Assembly Reassembly is the reverse of the above disassembly instructions, taking care that all sealing surfaces are clean and undamaged. 5-13 Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
  • Page 36: Safety Valve - Parts List

    P3000 Series Users Manual gjn37.eps Figure 5-7. Safety Valve Table 5-7. Safety Valve - Parts List Item Description Part Item Description Part Safety Valve 3919354 Screw 3908578 O Ring 3865142 Bonded Seal 3921964 Safety Plug 3919331 Vacuum Body 3921817 Plug Retainer...
  • Page 37: Fault Finding

    4. Ensure that the external supply is correctly connected and functioning correctly. 5. Ensure that the selector valve(s) are correctly positioned (if fitted). 6. Ensure that the safety valve (models with vacuum PCU only) is seated correctly. 7. Ensure that the DUT is not leaking. Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
  • Page 38: High Pcu Fall-Rate

    P3000 Series Users Manual High PCU Fall-Rate The piston will always fall slowly as the gas passes between the piston and cylinder. However, this fall rate will never be so fast that a stable reading cannot be made. 1. The system may be leaking. Check that exhaust valve is fully closed, and test port connections are correctly made.
  • Page 39: Storage & Transportation

    The weight set is heavy (can be up to 80 lbs/36 kg per individual box) so care must be taken when moving it. Both handles must be used when lifting the set for stability, and we recommend that it be carried between two people. Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
  • Page 40: Ancillary Equipment

    The top connection reproduces the test port on the deadweight tester, allowing the standard gauge adapters to be used. gjn030.bmp Figure 8-1. Dirt/Moisture Trap Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
  • Page 41: Angle Adapter, P5543

    P3000 Series Users Manual Angle Adapter, P5543 To calibrate gauges with the pressure connection on the rear (e.g. panel-mount gauges) in their correct position, an angle adapter should be used. The angle adapter fits directly onto the test station, converting it through 90 degrees, allowing the standard adapters to be used.