Fluke authorized resellers shall extend this warranty on new and unused products to end-user customers only but have no authority to extend a greater or different warranty on behalf of Fluke. Warranty support is available only if product is purchased through a Fluke authorized sales outlet or Buyer has paid the applicable international price.
Chapter Title Page 1 General Information ................1-1 Introduction......................1-1 How to Contact Fluke ..................1-1 Safety Information ..................... 1-1 Safety Summary .................... 1-1 Compressed Gas .................... 1-2 Heavy Weights ....................1-2 Personal Protective Equipment..............1-2 Symbols Used in this Manual ................1-2 Operating Principle ....................
Page 5
PCU Assembly - 16 mm - Parts List ..............5-7 5-4. PCU Assembly - 22 mm - Parts List ..............5-9 5-5. Handpump Assembly - Parts List................5-11 5-6. System Selector Valve - Parts List................. 5-13 5-7. Safety Valve - Parts List ..................5-14 Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
W Warning If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired. Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
AC (Alternating Current) Earth Ground Important Information: refer to manual Do not dispose of this product as unsorted municipal waste. Go to Fluke’s website for recycling information. Operating Principle Deadweight Testers are the primary standard for pressure measurement. Utilizing the...
Some countries have geographic/geological survey/mapping organizations that have the data readily available. If not, the countries’ National Standards Laboratory may be able to recommend a source of suitable information. Temperature Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
TRUE PRESSURE TRUE PRESSURE Model Variations This manual covers many pressure ranges and models within the P3000 series of instruments. The following illustrations detail the physical appearance of the six basic models: Model 3015 gjn024.bmp Figure 1-1. Single PCU Instrument Fluke-Direct .com...
General Information Model Variations Model 3012P, 3013P, 3014P and 3015P gjn025.bmp Figure 1-2. Single PCU Instrument with Handpump Model 3022, 3023 and 3025 gjn026.bmp Figure 1-3. Dual PCU (Pressure & Vacuum) Instrument Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
General Information Model Variations Model 3011P gjn029.bmp Figure 1-6. Vacuum Instrument with Handpump Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
Dirt/Moisture Trap, P5531, can be mounted to the test port of the deadweight tester. This will prevent any dirt particles or moisture droplets present within the DUT from passing into the deadweight tester system, (see Chapter 8). Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
P3000 Series Users Manual Connections Fit the device under test (DUT) to the test port using the method described below: W Warning DO NOT use Teflon/PTFE tape on these connections, as this will prevent correct sealing. The Gauge Adapter sealing system is designed for hand-tight sealing up to 20,000 psi / 1,400 bar-wrenches or similar tools are not required —...
Connections gjn007.bmp Figure 2-2. Connect Assembly to Test Port gjn008.bmp Figure 2-3. Hand-Tighten Only 3. To adjust the position to face forward, hold the gauge adapter and turn the instrument COUNTER-CLOCKWISE, so that it faces forward. Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
P3000 Series Users Manual gjn009.bmp Figure 2-4. Adjust Gauge Position 4. Hold the instrument steady, whilst turning the gauge adapter COUNTER-CLOCKWISE until it pulls down onto the O-ring. gjn010.bmp Figure 2-5. Tighten Gauge Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
Table 2-1. Test Port Inserts - Parts List Item Description Part O Ring 3865142 Test Port Insert 3919892 O Ring 3883397 Test Port 3921566 To calibrate panel-mounted gauges with pressure connections in the rear, use an Angle Adapter (see Chapter 8, Ancillary Equipment). Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
Chapter 3 Operation Introduction All P3000 Series pneumatic deadweight testers have a monitor gauge fitted to the instrument top plate, to serve as a guide to the operator by indicating the approximate pressure in the system. Single PCU Models — Pressure 1.
P3000 Series Users Manual 8. Depressurize the system by SLOWLY opening the Exhaust valve, as sudden depressurization will cause the weight stack to fall quickly, which may damage the piston assembly. 9. Remove the weights from the piston. Single PCU Models — Vacuum 1.
(e.g. local gravity, temperature, pressure head, etc.) to enhance the pressure measurement accuracy of the instrument. It will allow calibration in any of 12 different pressure units, using the existing weight set. Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
6. After removal from the solvent, using a NEW tissue, repeat the cleaning procedure in step 4. 7. Place piston carefully on a NEW tissue where it will not be damaged while the cylinder is cleaned. Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
P3000 Series Users Manual W Caution Never touch the working surface of a clean piston with bare fingers — the natural oil in your skin can cause the piston and cylinder to stick. 8. Wipe any dirt or moisture from the outside surfaces of the cylinder (2).
4. To remove all traces of contamination, the piston can be cleaned in a suitable solvent. W Caution O-ring seals (where fitted) are Nitrile rubber, and should not be immersed in solvents, as they will become damaged. They should be wiped carefully with a new tissue. Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
P3000 Series Users Manual 5. After removal from the solvent, using a NEW tissue, repeat the cleaning procedure in step 3, taking great care to ensure that all traces of solvent are removed from the inside of the tubular piston and weight carrier.
5. To remove all traces of contamination, the piston can be cleaned in a suitable solvent. W Caution O-ring seals (where fitted) are Nitrile rubber, and should not be immersed in solvents, as they will become damaged. They should be wiped carefully with a new tissue. Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
P3000 Series Users Manual 6. After removal from the solvent, using a NEW tissue, repeat the cleaning procedure in step 4, take great care to ensure that all traces of solvent are removed from inside the tubular piston. 7. Place piston carefully on a NEW tissue where it will not be damaged while the cylinder is cleaned.
5. To remove all traces of contamination, the piston can be cleaned in a suitable solvent. W Caution O-ring seals (where fitted) are Nitrile rubber, and should not be immersed in solvents, as they will become damaged. They should be wiped carefully with a new tissue. Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
P3000 Series Users Manual 6. After removal from the solvent, using a NEW tissue, repeat the cleaning procedure in step 4, take great care to ensure that all traces of solvent are removed from inside the tubular piston. 7. Place piston carefully on a NEW tissue where it will not be damaged while the cylinder is cleaned.
6. Carefully lift the top plate to reveal the inlet connection(s) from the external supply manifold (mounted on the rear panel of the case). 7. Disconnect the tubing, and lift the top plate clear of the case. Fluke-Direct .com info@Fluke-Direct.com...
P3000 Series Users Manual W Caution When handling the top plate assembly, it is good practice to remove the piston assemblies to avoid accidental damage. Note Replacement is simply the reverse of the above procedure. Handpump Assembly Disassembly 1. Disconnect tubing from check valve (19) and/or (23).
3918868 O Ring 3867548 Check Valve Body Spec O Ring 3918984 Coupling 3918991 Snap Ring 3918797 Flexible Tubing 3922387 Piston Shaft Spec. Coupling 3923564 O Ring 3883480 Check Valve Assy. 3920249 Piston Head Assy. Spec. 5-11 Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
P3000 Series Users Manual System Selector Valve Before commencing disassembly of the selector valve, make a note of the tubing connections to avoid problems during reassembly. Disassembly 1. Disconnect nylon tubing from connectors (11). 2. Loosen set screws (1), and remove knob (2) from valve shaft (3).
3. Withdraw safety valve (1) from safety plug (3), and inspect O ring (2) for cleanliness or signs of damage—replace as necessary. Re-Assembly Reassembly is the reverse of the above disassembly instructions, taking care that all sealing surfaces are clean and undamaged. 5-13 Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
P3000 Series Users Manual gjn37.eps Figure 5-7. Safety Valve Table 5-7. Safety Valve - Parts List Item Description Part Item Description Part Safety Valve 3919354 Screw 3908578 O Ring 3865142 Bonded Seal 3921964 Safety Plug 3919331 Vacuum Body 3921817 Plug Retainer...
4. Ensure that the external supply is correctly connected and functioning correctly. 5. Ensure that the selector valve(s) are correctly positioned (if fitted). 6. Ensure that the safety valve (models with vacuum PCU only) is seated correctly. 7. Ensure that the DUT is not leaking. Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
P3000 Series Users Manual High PCU Fall-Rate The piston will always fall slowly as the gas passes between the piston and cylinder. However, this fall rate will never be so fast that a stable reading cannot be made. 1. The system may be leaking. Check that exhaust valve is fully closed, and test port connections are correctly made.
The weight set is heavy (can be up to 80 lbs/36 kg per individual box) so care must be taken when moving it. Both handles must be used when lifting the set for stability, and we recommend that it be carried between two people. Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
The top connection reproduces the test port on the deadweight tester, allowing the standard gauge adapters to be used. gjn030.bmp Figure 8-1. Dirt/Moisture Trap Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
P3000 Series Users Manual Angle Adapter, P5543 To calibrate gauges with the pressure connection on the rear (e.g. panel-mount gauges) in their correct position, an angle adapter should be used. The angle adapter fits directly onto the test station, converting it through 90 degrees, allowing the standard adapters to be used.