Fluke CombiScope PM3370B User Manual page 286

Autoranging combiscope instrument
Hide thumbs Also See for CombiScope PM3370B:
Table of Contents

Advertisement

N - 2
B. Cable to printer/plotter with hardware handshake
parameters: "3-wire/7-wire" = 7-wire
C. Cable to controller with hardware handshake
parameters: "3-wire/7-wire" = 7-wire
The communication is confirm the RfR protocol. This means that signal RfR is
active when the oscilloscope can receive data.
NOTE: When during data transfer the signal DSR on the oscilloscope becomes
inactive, the serial communication buffers of the instrument are cleared.
The hard copy action or waveform transfer stops then.
"XON-XOFF on off" = on or off
OSCILLOSCOPE
(DTE)
SIGNAL
9 PIN
(female)
TxD
3
2
RxD
RfR
7
CTS
8
DTR
4
DSR
6
DCD
1
S.GND
5
F.GND
CASE
"XON-XOFF on off" = on or off
OSCILLOSCOPE
(DTE)
SIGNAL
9 PIN
(female)
TxD
3
RxD
2
RfR
7
CTS
8
DTR
4
DSR
6
DCD
1
S.GND
5
CASE
F.GND
RS-232 CABLE CONFIGURATIONS
PRINTER/PLOTTER
(DTE)
9 PIN
25 PIN
(male)
(male)
3
2
2
3
7
4
8
5
4
20
6
6
1
8
5
7
1
CASE
ST6921
CONTROLLER
(DTE)
9 PIN
25 PIN
(female)
(female)
3
2
2
3
7
4
8
5
4
20
6
6
1
8
5
7
1
CASE
ST6920

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents