THE CPL PROTOCOL
wave_nr
The oscilloscope waveform destination:
011 - 084 for m1.1 - m8.4
091 - 504 for m9.1 - m50.4 (Extended Memory only)
09e - 50e for m9.e - m50.e (Extended Memory only)
(xx.e for 2 ch. models only; xx.3 and xx.4 for 4 ch. models only)
admin
16 parameters, separated by a "pds":
PARAMETER
trace_name
Y_unit
X_unit
Y_zero
X_zero
Y_resolution
X_resolution
Y_range
date
time
dT-corr
min/max
mult_shot_tot
mult_shot_nr
reserved
reserved
Notes:
count
The number of samples in the waveform:
512, 1024, 2048, 4096, 8192, 16384 or 32768
(Extended Memory: 2048 and higher)
TYPE
EXAMPLE
string
m4.1
string
V
string
s
number
+3
number
-8.625E-6
number
78.13E-3
number
1E-6
number
65535
string
25-01-1994
string
17:10:25:40
number
375E-3
number
0
number
32
number
5
string
0
string
0
1) dT = dT-corr
*
2) 1 = min/max trace envelopes 0 = waveform trace
3) date format: European = E25-01-1994
American = A01-25-1994
Japanees = J1994-01-25
for compatibility reasons (initial versions)
European format without leading character can
be used.
4) for compatibility reasons (initial versions) hundreds of
seconds can be omited
01e - 08e for m1.e - m8.e
ACTION BY
OSCILLOSCOPE NOTES
ignored
interpreted
interpreted
interpreted
ignored
interpreted
interpreted
ignored
interpreted
interpreted
interpreted
interpreted
interpreted
interpreted
ignored
ignored
X-resolution
6 - 23
3)
4)
1)
2)

Hide quick links: