PROGRAM WAVEFORM
Purpose:
To send a waveform to the oscilloscope. This function is referred to as to Program
a complete Waveform in the oscilloscope.
A waveform is sent (programmed) in two command sequences. The first
sequence selects the waveform register number, programs the waveform
administration data, and programs the number of samples. The second sequence
programs the samples, including a checksum. Each sequence is synchronized by
an acknowledge response.
You can use the PW command to send the data received with the QW command
back to the oscilloscope.
Note:
The number of data bits of the RS-232 interface must be 8 (not 7),
otherwise, an execution error will occur.
Command 1:
"PW"
pds
Response 1:
acknowledge
Command 2:
sample
Response 2:
acknowledge
wave_nr
pds
CR
checksum
CR
THE CPL PROTOCOL
admin
pds
PW
count
CR