Fluke P5515-V User Manual
Fluke P5515-V User Manual

Fluke P5515-V User Manual

Hydraulic comparison test pump

Advertisement

Quick Links

P5515, P5515-V, &
P5515-EP
Hydraulic Comparison Test Pump
Users Manual
PN 3952251
November 2010
© 2010 Fluke Corporation. All rights reserved. Printed in USA. Specifications are subject to change without notice.
All product names are trademarks of their respective companies.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the P5515-V and is the answer not in the manual?

Questions and answers

Summary of Contents for Fluke P5515-V

  • Page 1 P5515, P5515-V, & P5515-EP Hydraulic Comparison Test Pump Users Manual PN 3952251 November 2010 © 2010 Fluke Corporation. All rights reserved. Printed in USA. Specifications are subject to change without notice. All product names are trademarks of their respective companies.
  • Page 2 To register your product online, visit register.fluke.com...
  • Page 3: Table Of Contents

    Table of Contents Title Page Introduction......................1 How to Contact Fluke ..................2 Safety Information ..................... 2 Safety Summary .................... 2 Compressed Liquid..................2 Personal Protective Equipment..............2 Symbols Used in this Manual ................2 Preparation ......................3 Operating Fluid Compatibility................3 Connections .......................
  • Page 4 P5515, P5515-V, & P5515-EP Users Manual...
  • Page 5 List of Tables Table Title Page Symbols........................3 Increased Fluid Compatibility Models..............3 Test Port Insert - Parts.................... 7 Screw Pump Assembly - Parts List ................ 10 Priming Pump Assembly - Parts List ..............13 Check Valve Assemblies - Parts List ..............14 Reservoir Assembly - Parts List................
  • Page 6 P5515, P5515-V, & P5515-EP Users Manual...
  • Page 7 List of Figures Figure Title Page Hydraulic Circuit Schematic .................. 1 Screw on Gauge Adapter..................4 Mount Gauge......................5 Adjust Gauge Position.................... 5 Tighten Gauge ......................6 Do Not Over-Tighten ..................... 6 Test Port Insert ....................... 7 Screw Pump Assembly ..................10 Priming Pump Assembly..................
  • Page 8 P5515, P5515-V, & P5515-EP Users Manual...
  • Page 9: Introduction

    Introduction This manual contains operation and routine and preventive maintenance instructions for the Model P5515, P5515-V, and P5515-EP Hydraulic Comparison Test Pump. The manual will refer to all model numbers of the instrument listed above generically as a Model P5515, unless denoted as only applicable to a specific model number.
  • Page 10: How To Contact Fluke

    P5515, P5515-V, & P5515-EP Users Manual How to Contact Fluke To order accessories, receive operating assistance, or get the location of the nearest Fluke distributor or Service Center, call: Technical Support USA: 1-800-99-FLUKE (1-800-993-5853) Calibration/Repair USA: 1-888-99-FLUKE (1-888-993-5853) Canada: 1-800-36-FLUKE (1-800-363-5853)
  • Page 11: Preparation

    Earth Ground Important Information: refer to manual Do not dispose of this product as unsorted municipal waste. Go to Fluke’s website for recycling information. Preparation The comparison test pump must be set up on a level, stable workbench or similar surface.
  • Page 12 P5515, P5515-V, & P5515-EP Users Manual To avoid cross-contamination from other fluids, and protect the system from particulates, we recommend the use of a Liquid Separator, (refer to section Ancillary Equipment). Warning DO NOT use Teflon/PTFE tape on these connections, as this will prevent correct sealing.
  • Page 13 Hydraulic Comparison Test Pump Connections gkx03.bmp Figure 3. Mount Gauge gkx05.bmp Figure 4. Adjust Gauge Position 3. To adjust the position to face forward, hold the gauge adapter and turn the instrument COUNTER-CLOCKWISE, so that it faces forward.
  • Page 14 P5515, P5515-V, & P5515-EP Users Manual gkx06.bmp Figure 5. Tighten Gauge 4. Hold the instrument steady, while turning the gauge adapter COUNTER-CLOCKWISE until it pulls down onto the O-ring. gkx07.bmp Figure 6. Do Not Over-Tighten...
  • Page 15: Test Port Insert

    (stored in the spare seals container) as shown in Figure 7. gjn012.eps Figure 7. Test Port Insert Table 3. Test Port Insert - Parts Part Item Description P5515 P5515-V P5515-EP O Ring 3865142 3865163 3867888 Test Port Insert 3919892...
  • Page 16: Operation

    The part number for the seal kit for the P5515, P5515-V, and P5515-EP is P5515SK, P5515SK-V, and P5515SK-EP, respectively Top Plate Removal 1.
  • Page 17: Screw Pump Assembly

    Hydraulic Comparison Test Pump Maintenance and Servicing 3. Remove the spokes from the hub on the front of the instrument. 4. Remove the 4 screws from the instrument top plate, (1 at the mid-point of each edge). 5. Hold the top plate assembly by the test port, and tilt the plate so that the rear edge is lifted, but the front edge remains in contact with the instrument case.
  • Page 18 P5515, P5515-V, & P5515-EP Users Manual gkx09.eps Figure 8. Screw Pump Assembly Table 4. Screw Pump Assembly - Parts List Part Item Description P5515 P5515-V P5515-EP Test Port 3921414 O Ring 3864922 3865332 3865565 Barrel Locknut 3921438 Barrel 3921450 Barrel Support...
  • Page 19: Priming Pump Assembly

    Hydraulic Comparison Test Pump Maintenance and Servicing Priming Pump Assembly Disassembly 1. On the underside of the top plate, disconnect the hydraulic tubing from the pump block assembly (14). 2. On the upper side of the top plate, unscrew locknut (3) ½ turn, and remove clip pin from clevis (2).
  • Page 20 P5515, P5515-V, & P5515-EP Users Manual gkx10.eps Figure 9. Priming Pump Assembly...
  • Page 21: Check Valves

    Hydraulic Comparison Test Pump Maintenance and Servicing Table 5. Priming Pump Assembly - Parts List Part Item Description P5515 P5515-V P5515-EP Pivot Clip 3920224 Clevis 3920236 Locknut 3918713 Pump Union Nut 3921527 Shaft Bearing 3867744 Pump Shaft 3921509 Ball 3917855...
  • Page 22: Reservoir Assembly

    P5515, P5515-V, & P5515-EP Users Manual gkx11.eps Figure 10. Check Valve Assemblies Table 6. Check Valve Assemblies - Parts List Part Item Description P5515 P5515-V P5515-EP Pump Block 3921530 O Ring 3864766 O Ring 3867636 3867912 3867874 Bullet 3920584 Spring...
  • Page 23 Hydraulic Comparison Test Pump Maintenance and Servicing 5. To completely disassemble the reservoir, the top plate must be removed from the instrument case. 6. Disconnect the high-pressure tubing by unscrewing the gland nuts fitted to the reservoir body (12). 7. The low pressure feed to the priming pump is disconnected by pressing down on the black plastic ring at the top of the connector, and gently pulling on the nylon tubing.
  • Page 24: Fault Finding

    P5515, P5515-V, & P5515-EP Users Manual Table 7. Reservoir Assembly - Parts List Part Item Description P5515 P5515-V P5515-EP VALVE STEM 3886430 SPRING 3922786 NYLON WASHER 3916458 RESERVOIR COVER 3921624 RESERVOIR CAP 3921613 O RING 3864910 3865326 3865552 RESERVOIR TUBE...
  • Page 25: Priming Pump Malfunction

    Hydraulic Comparison Test Pump Storage and Transportation Priming Pump Malfunction 1. If pumping does not generate pressure, then the inlet check valve has probably failed, and requires replacement. 2. If the system pressurizes and depressurizes in conjunction with the downward and upward strokes of the pump, then the outlet check valve has failed completely.
  • Page 26: Ancillary Equipment

    P5515, P5515-V, & P5515-EP Users Manual Ancillary Equipment Angle Adapter, P5543 To calibrate gauges with the pressure connection on the rear (e.g. panel-mount gauges) in their correct position, an angle adapter should be used. The angle adapter fits directly onto the test station, converting it through 90 degrees, allowing the standard adapters to be used.
  • Page 27: Liquid Separator, P5521

    Hydraulic Comparison Test Pump Ancillary Equipment Liquid Separator, P5521 The Liquid Separator is used to separate the working fluid of the deadweight tester to which it is fitted from the working fluid in the device under test. This prevents contamination of either hydraulic system and allows calibration of the device in its specific working fluid.
  • Page 28 P5515, P5515-V, & P5515-EP Users Manual...