Setup and Operation

Once the analyzer has been installed, it can be configured to your
application. To do this you will:
• Set system parameters
• Calibrate the instrument
• Define the analysis range. Then choose auto-ranging or
select a fixed range of analysis, as required.
• Set alarm setpoints, and modes of alarm operation (latching,
failsafe, etc.)

4.1 Analyzer Startup

Referring to the Installation section and "AS-BUILT" Drawings in
the Appendix, check the following:
1. Air/oxygen supply – This must provide a constant
pressure of 15 +/- 2.0 psi, ultra-pure oxygen or CO
air at a flow rate up to 500 cc/minute. (Pre-purified air
may be used with optional Air Purifier P/N ST910003).
2. Drain Line – Check to see if the drain line is free of kinks
and loops and allows normal vented gravity flow to an
open receptacle (jar, pipe, etc.). AN AIR GAP AS
DEFINED IS MANDATORY.
CAUTION: FAILURE TO OBSERVE THIS WILL RESULT IN
IMPROPER OPERATION.
3. ELECTRICAL – Assure proper electrical installation.
Refer to facility requirements.
4. SAFETY – All customer and Teledyne specified safety
measures are followed.