Theory Of Operation; Overview; Background - Teledyne 6750 Operating Instructions Manual

Total organic carbon analyzer
Table of Contents

Advertisement

Total Organic Carbon Analyzer

Theory of Operation

2.1 Overview

The Model 6750 Analyzer uses the UV/Heated Persulfate method
of analysis and is well-suited for many applications involving accurate
TOC analysis. The basic analyzer is configured for maximum utility
using an advanced Microsoft
screen. The analyzer is suitable for both general purpose and/or
hazardous area classifications, if properly configured with required
safety equipment.

2.2 Background

TOC analysis is the primary screening tool and control parameter
for all water based applications but there are some discrepancies or
misunderstanding in just what TOC is comprised of.
Many manufacturers claim TOC analysis capability without
disclosing the limitations of their methods. Some treat all "TOC
analyzers" as a commodity, with no distinguishing characteristics among
them. Analytical results reported by the user can therefore be
questionable, unless the TOC method is matched to the analytical
requirement.
The following discussion is intended to provide the user adequate
information in order to allow intelligent cost/performance trade-offs and
select the appropriate TOC method for each application. The Model 6750
TOC Analyzer performs the following analyses on the SAME sample:
• TC:
• NPOC:
• TIC:
• POC (VOC): Purgeable Organic Carbon or Volatile Organic
• TOC-True :
Teledyne Analytical Instruments
Windows-based CE computer with touch
Total Carbon
Non-Purgeable Organic Carbon
Total Inorganic Carbon
Carbon
Total Organic Carbon
Theory of Operation
5

Advertisement

Table of Contents
loading

Table of Contents