Total Organic Carbon Analyzer
be reduced to approximately 2.0 to 3.0. If the sample pH is not properly
reduced as described in Figure 2-8, increase the normalcy of the acid
reagent until the water exiting the sparger drain is between 2.0 and 3.0.

Figure 2-8: Relationship Between TIC and pH

2.4.3 Oxidation

The UV reactor and persulfate reagent provide oxidation of the
carbon to CO
. The UV reactor consists of the following:
2
• Mounting enclosure with viewing window
• Borosilicate reactor body
• Heater/Temperature sensor
• UV Lamp
• Miscellaneous fittings and clamps
Teledyne Analytical Instruments
Theory of Operation
15