Figure 2-1: Npoc Or Toc - Direct Method; Figure 2-2: Toc - True Method - Teledyne 6750 Operating Instructions Manual

Total organic carbon analyzer
Table of Contents

Advertisement

Theory of Operation
To perform correct TOC analysis, the analysis system must be
capable of measuring all constituents of organic carbon present in the
sample: Non-Purgeable Organic Carbon (NPOC) and Purgeable
Organic Carbon (POC). Correct TOC analysis must also exclude the
Total Inorganic Carbon (TIC) interference.
Figure 2-1 is a basic, commonly used technique to measure the
NPOC, sometimes referred to as "TOC-Direct".
Figure 2-1: NPOC or TOC – Direct Method
In this method, acid is added to the sample, lowering its pH to
approximately 2.0, at which point the carbonates present in the sample are
converted to dissolved CO
the CO
converted from the TIC and vents it, along with any purgeable
2
(volatile) organics, leaving only NPOC in the sample. The resultant NPOC
is then oxidized to CO
NPOC in the sample, often referred to and reported erroneously as "TOC".
In Figure 2-2 a preferred method of performing a "TOC- True"
analysis is shown.
In this case, acid is added to the sample, lowering its pH to
approximately 2.0, at which point the carbonates present in the sample
are converted to dissolved CO
Teledyne Analytical Instruments
. In the sparger, the carrier gas strips (sparges)
2
in the reactor and measured by the CO
2
Figure 2-2: TOC – True Method
. In the sparger, the carrier gas strips
2
Model 6750
detector as
2
6

Advertisement

Table of Contents
loading

Table of Contents