Total Organic Carbon Analyzer
All operator controls and serviceable components are mounted on
the control panel, which is hinged and doubles as the door that provides
access to those components not normally requiring maintenance.
Internal components of the Model 6750 TOC Analyzer are shown
in Figure 3-3.

Figure 3-3: Internal Components of the Model 6750 Analyzer

3.3 Electrical Connections

Figure 3-4 shows the analyzer electrical/electronics connections for
power, communication, and both digital and analog concentration
outputs. See also the AS-BUILT drawing package included in the
Appendix.

3.3.1 Power Connection

The universal power supply requires an 85-250 VAC, 47-63 Hz
power source.
For safe connections, ensure that no uninsulated wire extends
outside the connectors they are attached to. Stripped wire ends must
insert completely into terminal blocks. No uninsulated wiring should be
able to come in contact with fingers, tools or clothing during normal
operation. Electrical/Electronics hookups include the following:
Teledyne Analytical Instruments
Installation
23