8.4.4.8
8.4.4.9
8.4.4.10
98
CALC2:CONV:PAR:VAL RTPW, 100.0145, A4, 0.0045
The CALCulate suffix, <chn>, specifies the channel (1 to 4) or memory chan-
nel (5 to 14). The <param> parameter is the name of the parameter as given in
Table 8, Conversion Types and Parameters. If the <param> parameter is not ap-
plicable to the current conversion type, an error -221, "Settings Conflict", is
generated. The <num> parameter is the desired value of the parameter. Multiple
parameters can be set using a single command by separating the <param> and
<num> prameters by commas. The order of the parameters is not important but
the value must immediately follow the parameter name.
This command may be password-protected. To access this command:
If the SYST:PASS:CONV? command returns a 0 (probe password-protec-
tion is disabled), the password is not required.
If the SYST:PASS:CONV? command returns a 1 (probe password-protec-
tion is enabled), the password must be successfully entered using the
SYST:PASS:CEN <pass> command (see Section 9.4.11, Password Com-
mands).
CALCulate:CONVert:PRINt [<chn>|ALL [,<port>]]
Prints the conversion parameters for the specified channel.
CALC:CONV:PRIN 2
"PRB",CHAN2,"ser",090105,"TYP",RES,"RANGE",0
If no parameters are specified, the ALL parameter is assumed and the parame-
ters for all channels are printed to the current port. Specifying the <chn> pa-
rameter prints only the parameters for the specified channel (1 to 4) to the
current port. Optionally, the parameters may be printed to a specific port by
specifying the <port> parameter (see Table 10 on page 129, Port Numbers)
CALCulate<chn>:CONVert:SNUMber?
Returns the probe serial number for the specified channel.
CALC2:CONV:SNUM?
A_336C
The CALCulate suffix, <chn>, specifies the channel (1 to 4) or memory chan-
nel (5 to 14). The returned serial number is in string format.
CALCulate<chn>:CONVert:SNUMber <serl>
Sets the probe serial number for the specified channel.
CALC2:CONV:SNUM A_336C
The CALCulate suffix, <chn>, specifies the channel (1 to 4) or memory chan-
nel (5 to 14). The <serl> parameter is in string format. The <serl> parameter