Logging:automatic:print [<Num>|All [,<Port>]]; Logging:automatic:status; Logging:automatic:status <Bool; Logging:automatic:time? [Min|Max|Def] - Fluke 1529 Chub-E4 User Manual

Chub-e4 thermometer readout
Table of Contents

Advertisement

1529 Chub-E4 Thermometer Readout
User's Guide
8.4.7.8
8.4.7.9
8.4.7.10
8.4.7.11
108
LOGging:AUTomatic:PRINt [<num>|ALL [,<port>]]
Prints the auto log entries.
LOG:AUT:PRIN ALL
DATA_25 1 22.676C 12:19:42 09-05-00
DATA_25 2 9.960 KO 12:19:44 09-05-00
DATA_25 3 23.220C 12:19:46 09-05-00
DATA_25 4 23.245C 12:19:48 09-05-00
DATA_25 1 22.765C 12:19:52 09-05-00
DATA_25 2 9.962 KO 12:19:54 09-05-00
DATA_25 3 23.087C 12:19:56 09-05-00
DATA_25 4 23.260C 12:19:58 09-05-00...
If no parameters are specified, the ALL parameter is assumed and all auto log
entries are printed to the serial port. Specifying the <num> parameter prints
only the log entries with the specified label number to the serial port. Op-
tionally, the log entries may be printed to a specific port by specifying the
<port> parameter (see Table 10 on page 129, Port Numbers).

LOGging:AUTomatic:STATus?

Returns the auto log status.
LOG:AUT:STAT?
0
If this query returns a 0, an auto log session is not running. If this query returns
a 1, an auto log session is running.
LOGging:AUTomatic:STATus <bool>
Sets the auto log status.
LOG:AUT:STAT 0
The <bool> parameter turns auto logging on (1 or ON) or off (0 or OFF). If si-
multaneous scan mode is enabled, one measurement is stored for each channel
in the auto log. The *RST command set the auto log status to off.

LOGging:AUTomatic:TIMe? [MIN|MAX|DEF]

Returns the auto log interval setting.
LOG:AUT:TIM?
1
Specifying the MIN parameter returns a value of 0.1. Specifying the MAX pa-
rameter value returns value of 3600. Specifying the DEF parameter returns a
value of 1. The units are seconds.