Format:stamp; Format:stamp <Bool; Measure? [<Chn>]; Read? [<Chn>] - Fluke 1529 Chub-E4 User Manual

Chub-e4 thermometer readout
Table of Contents

Advertisement

1529 Chub-E4 Thermometer Readout
User's Guide
8.4.1.5
8.4.1.6
8.4.1.7
8.4.1.8
8.4.1.9
90
day the measurement was made
With this format kohms and millivolts are not converted to ohms and volts.
This command may return the same reading if a new reading has not yet been
obtained. If only new readings are desired, the FORMat:STAMp command can
be used to enable the extended format option and any readings in which the
value of the first field is 0 can be ignored or else the STATus:MEASure? com-
mand can be used to determine if a new reading is available for a desired
channel.

FORMat:STAMp?

Returns the extended measurement data format setting. See Section 9.4.1.4,
FETCh? for details.
FORM:STAM?
0
FORMat:STAMp <bool>
Enables extended measurement data format with the FETCh?, MEASure? And
READ? commands. On power up this is always set to off.
FORM:STAM ON
MEASure? [<chn>]
Returns the most recent measurement for the specified channel.
MEAS? 1
0.0127
This command is the same as the FETCh? [<chn>] command
READ? [<chn>]
Returns the most recent measurement for the specified channel.
READ? 2
0.0113
This command is the same as the ?FETCh [<chn>] command.
SENSe<chn>:DATA?
Returns the input value for the specified channel.
SENS3:DATA?
100.0291, 0.0000

Hide quick links:

Advertisement

Table of Contents

Troubleshooting

loading

This manual is also suitable for:

1529

Table of Contents