Calibrate<Chn>: Parameter:linearity<N> <Num>|Min|Max|Def; Calibrate<Chn>:Parameter:rjc? [Min|Max|Def]; Calibrate<Chn>:Parameter:rjc <Num>|Min|Max|Def - Fluke 1529 Chub-E4 User Manual

Chub-e4 thermometer readout
Table of Contents

Advertisement

suffix <n> is 1 (for PRTs) and 9000 when the linearity suffix <n> is 2 (for
Thermistors). Specifying the DEF parameter returns a value of 0.
8.4.5.11
CALibrate<chn>: PARameter:LINearity<n> <num>|MIN|MAX|DEF
Sets the calibration linearity parameter for the specified channel.
CAL1:PAR:LIN1 5
This command only applies to PRT/Thermistor channels. Executing this com-
mand on Thermocouple channels generates a -294, "Incompatible type" error
message.
The CALibrate suffix, <chn>, specifies the channel (1 to 4). This command
does not apply to memory channels. For PRT/Thermistor channels, the calibra-
tion scale suffix <n> specifies the resistance range parameter (in ohms) for
PRTs (1) or Thermistors (2). The <num> parameter specifies the value for the
specified linearity parameter. Specifying the MIN, MAX, or DEF parameter
sets the linearity parameter to the minimum, maximum, or default value
respectively.
This command is password-protected. To access this command, the password
must be successfully entered using the SYST:PASS:CEN <password> com-
mand (see Section 9.4.11, Password Commands).
8.4.5.12
CALibrate<chn>:PARameter:RJC? [MIN|MAX|DEF]
Returns the calibration RJC offset parmeter for the specified channel.
CAL1:PAR:RJC?
2.1
This command only applies to Thermocouple channels. Executing this com-
mand on PRT/Thermistor channels generates a -294, "Incompatible type" error
message.
The RJC suffix, <chn>, specifies the channel (1 to 4). This command does not
apply to memory channels. Specifying the MIN parameter returns a value of
–9. Specifying the MAX parameter returns a value of 9. Specifying the DEF
parameter returns a value of 0.
8.4.5.13
CALibrate<chn>:PARameter:RJC <num>|MIN|MAX|DEF
Sets the calibration RJC offset parameter for the specified channel.
CAL3:PAR:RJC 1.6
This command only applies to Thermocouple channels. Executing this com-
mand on PRT/Thermistor channels generates a -294, "Incompatible type" error
message.
The RJC suffix, <chn>, specifies the channel (1 to 4). This command does not
apply to memory channels. The <num> parameter specifies the RJC value for
8 Digital Communications Interface