Self-Test Error Messages; Start-Up Error Messages - Fluke 1529 Chub-E4 User Manual

Chub-e4 thermometer readout
Table of Contents

Advertisement

Message
SYSTEMS: IR PORT FAILURE
SYSTEMS: BATTERY NOT
INSTALLED
CHANNELS: FAILURE
CALIBRATION: DUE FOR RECAL
CALIBRATION: INVALID
PARAMETERS
GPIB: NOT INSTALLED
MEMORY DATA: QUESTIONABLE
MEMORY DATA: INVALID DATA
BUTTONS: [button name]
11.1.2
Message
CALIBRATION EXPIRED
INVALID CALIBRATION PARAMETERS
MEMORY QUESTIONABLE
MEMORY DATA INVALID, MEMORY WILL BE
CLEARED
11.2
Recommended Action
Cycle the power.
Cycle the power.
Check that the battery is properly connected.
Replace the battery.
Cycle the power.
Check that the instrument date is correct.
Recalibrate the instrument and set the recalibration date accordingly.
Check the correctness of the calibration parameters for each channel.
Normal message if the IEEE-488 option is not installed.
Reset the memory by pressing EXIT when prompted.
Reset the memory by pressing ENTER when prompted. (This action requires that important data
such as probe coefficients be restored.
Check the stated button to ensure that it is not stuck. Loosen as necessary.

Start-up Error Messages

The following messages may appear after the self-test. If the recommended ac-
tion fails to correct the error, contact a Hart Authorized Service Center (see
Section 1.3) for assistance.
Recommended Action
Check that the instrument date is correct.
Recalibrate the instrument and set the recalibration date accordingly.
Check the correctness of the calibration parameters for each channel.
Reset the memory by pressing EXIT when prompted.
Reset the memory by pressing ENTER. (This action requires that important data
such as probe coefficients be restored.)
Downloading Auto Logged Data
The 1529 "Chub-E4" Thermometer Readout stores up to 8,160 autolog read-
ings or 100 demand log readings in its memory. These readings can be down-
loaded to a computer using the RS-232 serial port.
The following steps should be followed to download logged data using the se-
rial port:
1. Connect the 1529 to a COM port on the computer using the serial cable.
2. Run a terminal program such as Windows® Terminal or HyperTerminal®.
11 Troubleshooting
Downloading Auto Logged Data
141