Logging:demand:store; Logging:demand:value? <Num>|Min|Max|Def; Logging:label<N>:Name; Logging:label<N>:Name <Label - Fluke 1529 Chub-E4 User Manual

Chub-e4 thermometer readout
Table of Contents

Advertisement

8.4.7.20

LOGging:DEMand:STORe

Stores the current measurement into the demand log.
LOG:DEM:STOR
If simultaneous scan mode is enabled, one reading is stored in the log for each
enabled channel.
8.4.7.21
LOGging:DEMand:VALue? <num>|MIN|MAX|DEF
Returns the specified demand log entry.
LOG:DEM:VAL? 10
DATA_25,2,9.961,KO,14,59,35,2000,9,5
The <num> parameter specifies the demand log entry to return. Specifying ei-
ther the MIN or DEF parameter returns the first entry in the demand log. Spec-
ifying the MAX parameter returns the last entry in the demand log. Each entry
in the demand log consist of the label, channel number, measurement value,
scale, time and date.
8.4.7.22
LOGging:LABel<n>:NAME?
Returns the name of the specified label.
LOG:LAB18:NAME?
DATA_18
The label suffix, <n>, specifies the number of the label (1 to 25) to retrieve.
The label name is returned in string format.
8.4.7.23
LOGging:LABel<n>:NAME <label>
Sets the name of the specified label.
LOG:LAB18:NAME BATH
The label suffix, <n>, specifies the number of the label (1 to 25) to set. Label
names can be up to 8 characters in length and can include any letters, numeric
digits, and the underscore ('_').
8.4.8

System Commands

The system commands are used for querying instrument identification parame-
ters, temperature units, and for setting operating parameters to the defined de-
fault values.
8.4.8.1

*IDN?

Returns the instrument identification string that indicates the manufacturer,
model number, serial number, and code firmware version.
8 Digital Communications Interface
Commands
111

Hide quick links:

Advertisement

Table of Contents

Troubleshooting

loading

This manual is also suitable for:

1529

Table of Contents