Directions for Use
by swabbing or wiping the surface with one of
the solutions mentioned above. Immerse the
probe well in mild detergent solution as
necessary for cleaning.
DO NOT use hard or sharp objects to clean
the probe well. This could damage the
probe well and cause the unit to not
function properly.
DO NOT use steam, heat, or gas
sterilization on the probe well.
DO NOT autoclave the probe well.
• Thoroughly dry all surfaces before re-
assembling the instrument.
• Re-connect the latching probe connector to the
thermometer. Ensure that the connector snaps
into place.
• Re-install the probe well in the thermometer and
snap the probe well into place.
• Insert the probe into the probe well.

Disposal

Disposing of Equipment

This equipment contains no hazardous materials.
Discard without environmental risk. Remove the
batteries before disposal.

Disposing of Batteries

Dispose of used batteries properly. Consult your
local recycling authority for proper procedures.
Performance and Technical
Specifications
Ambient
50.0 °F to 104.0 °F (10.0 °C
to 40.0°C)
Operating
Temperature
Range
Operating Altitude
-557 ft. to +16,000 ft / -170m
to +4877m
80.0 °F to 110.0 °F (26.7 °C
Patient
to 43.3 °C)
Temperature
Range
Temperature
Oral:
Predict Time*
Approximately 4–6 secs.
Adult axillary (18 years and
older):
Approximately 12–15 secs.
Pediatric axillary (17 years
and younger):
Approximately 10–13 secs.
Rectal:
Approximately 10–13 secs.
Transport/Storage
-13°F to +131°F (-25°C to
+55°C)
Temperature
Calibration
+/- 0.2 °F (0.1 °C) (Monitor
Mode)
Accuracy
Humidity
15% to 95% non-condensing
Dimensions
8.46" x 3.18" x 2.43"
(215mm x 81mm x 62mm)
Weight
12.6 ounces/357 grams
Power
(3) 1.5Vdc AA batteries
Equipment
Class 2 Device, Continuous
Operation
Classification
Not AP or APG equipment
IPXØ
Not protected against the
ingress of water.
Predict time depends on probe placement and
*
patient condition. Predict times do not include tissue
contact-detection time or time needed to place the
probe into the measurement site.

Replacement Parts and Accessories

To order replacement parts and accessories,
please contact the Welch Allyn office nearest to
you.

Service

Except for the replaceable AA batteries, this
instrument contains no user-serviceable parts.
Return the thermometer to a Welch Allyn
authorized service center for service and repair.
Refer to the Technical Manual.

Welch Allyn Service Policy

Only a Welch Allyn Service Center can perform or
approve all repairs on products under warranty.
Unauthorized repairs will void the warranty.
Qualified electronics personnel or at a Welch Allyn
Service Center should repair out-of-warranty
products.