Additional Specifications

The following paragraphs provide additional specifications for the 5500A Calibrator ac
voltage and ac current functions. These specifications are valid after allowing a warm-up
period of 30 minutes, or twice the time the 5500A has been turned off. All extended
range specifications are based on performing the internal zero-cal function at weekly
intervals, or when the ambient temperature changes by more than 5°C. (See Chapter 4,
"Front Panel Operations.")

Frequency Specifications

Frequency
Range
.01 to 119.99 Hz
120.0 to 1199.9 Hz
1.200 to 11.999 kHz
12.00 to 119.99 kHz
120.0 to 1199.9 kHz
1.200 to 2.000 MHz
[1] ± (25 ppm + 15 mHz) above 10 kHz
Resolution
1-Year Absolute Uncertainty,
.01 Hz
0.1 Hz
1.0 Hz
10 Hz
100 Hz
1 kHz
tcal + 5°C
25 ppm, + 1 mHz
25 ppm, + 1 mHz
25 ppm, + 1 mHz [1]
25 ppm, + 15 mHz
25 ppm, + 15 mHz
25 ppm, + 15 mHz
Multi-Product Calibrator
Additional Specifications
Jitter
2 µs
2 µs
2 µs
140 ns
140 ns
140 ns
29