Download Print this page

NSD Absocoder NCW-3DHPRV2 Specifications & Instruction Manual page 38

For iron and steel industry, converter for profibus-dp, applicable sensor vre-16ts100

Advertisement

Error
Description
Name
PROFIBUS-DP power supply error
(Error is "1")
Problem exists at converter's
PRFPF
internal power supply.
As communication is disabled,
check the PRFPF lamp at the LED
display.
Memory error (Error is "1")
ME
A data memory error has occurred.
Watchdog timer error (Error is "1")
WDTE
Runaway condition at the internal
CPU.
Not Ready (Error is "1" or operation
not yet started; Normal is "0")
NRDY
One of PRFPF, ME, or WDTE error
has occurred.
The internal power supply error for
sensors (Error is "1")
SPF1
SPF2
The power supply inside of the
converter is broken down.
Disconnected sensor error (Error is
"1")
SSE1
SSE2
The sensor of a correspondence
axis is not connected normally.
Sensor error (Error is "1")
SE1
SE2
Either SPF or SSE has occurred.
When
Status
Detected
This converter is not
Always
recognized by
PROFIBUS-DP.
An undetermined
At power
position data status
ON
exists at both axes.
An undetermined
Always
position data status
exists at both axes.
An undetermined
Always
position data status
exists at both axes.
An undetermined
Always
position data status
exists at both axes.
The position data of the
Always
axis which has an error
is "undetermined data".
Always
32
CLEAR
Countermeasure
Method
If there are no problems with
the PROFIBUS-DP cable or
CLR
communication, the
converter has failed.
CLR
Converter failure.
Restart
Converter failure.
Comply with
each error
clear
method.
CLR
Converter failure.
- Sensor connector is loose.
- Sensor cable is severed.
Automatic
- ABSOCODER sensor
recovery
failure
- Converter failure.
Comply with
each error
clear
method.

Advertisement

loading
Need help?

Need help?

Do you have a question about the Absocoder NCW-3DHPRV2 and is the answer not in the manual?

Subscribe to Our Youtube Channel