Shenchen Peristaltic Pump
E、If request higher accuracy, you can click Add and Dec button to micro adjust the
flow rate, to reach high accuracy transferring and dispensing.
F、Click Cancel button, cancel the test data, the actual volume return to 0.
Online Micro Adjust Volume Process:
Flow Rate Transferring Mode: If the actual flow rate is big or small than the set
up flow rate, you can micro adjust the flow rate online without affect the product
line.
Fixed Time and Volume Mode: If the dispensing volume is big or small than the
set up volume, you can micro adjust the volume online, no need to stop the pump.
Fixed Volume Measurement Mode: Do not support online micro adjust function.
①.Click the Calibration button from the main interface, enter the flow rate
calibration interface.
②. Now only the Add, Dec and Esc button is usable, other buttons are forbidden.
③. Click Add or Dec button to adjust the flow rate or volume.
10、Date & Time Interface:
Setting Date & Time Interface as below:
Set Date
Set Time
Click the System Setting button from the main interface, click Date and Time
button, enter date and time setting interface. The date and time will display on
12-hour
24-hour
- 11 -
2012-3-15
8:45:35 AM
Thursday
Back
LabV Series