Shenchen Precision Pump LabV6 Manual page 13

Peristaltic pump
Hide thumbs Also See for LabV6:

Advertisement

Shenchen Peristaltic Pump
FixedVolume
Volume
10.00
Run Time
1.00
1.00
The top left corner display the function, when fixed volume measurement turn
on, display fixed volume; when fixed time and volume turn on, display fixed
time and volume. Others display transferring mode.
If fixed time and volume turn on, the target volume and running time is set up
parameter, unable to amend. Other modes, the running time is 60s, you can
click the run time button to amend the running time.
Before the pump working, need to calibrate the flow rate to ensure the
transferring or dispensing accuracy
Process as below:
A、Confirm the running time, if fixed time and volume function, the running time is
set up time, unable to change.
B、Click Start button to test, countdown display the run time, it will stop
automatically, and display numerical keyboard, input the actual volume, then it
will ask whether continue test ( suggest more than 3 times), choose Yes, the
pump will test again, choose No, back to the calibration interface.
C、After click the Start button, during the pump running, you can click the Stop
button to stop the test.
D、After finish the tests, the actual volume area display the average data, click the
CAL button, the calibration is finished. Now the request flow rate or volume is
close to theoretical data.
Actual Vol.
0.0000ml
Test
ml
CAL
Reset
s
- 10 -
LabV Series
Volume adjust
+0.0000ml
Add
Dec
Esc