Summary of Contents for Shenchen Precision Pump N6 Series
Page 1
Shenchen Precision Pump Manual of N6 Series...
Page 2
Warning Tubing may have crack due to wear. It results in the overflow of fluid from tubing. In that time human body and instruments may be hurt. Must check usually and change tubing in time. Please connect directly the power line to the wall socket, and avoid ...
Shenchen Pump N6 Series 1. N6 Series Product Introduction N6 series products with 303 stainless steel housing, 3.2 inch LCD screen display; support variety of external control, RS232/RS485 communication, standard MODBUS protocol (RTU mode), meet different requirements. Suitable head: DZ25-3L, DZ25-6L, YZ35 2.
Shenchen Pump N6 Series 3. N6 Series Keypad Instruction Start/stop button: Press this button, the pump start running. Press this button again, pump stop running. Change direction button: Press this button to change the pump running direction. Full speed button: When the pump is stop or in transferring state, press this button, pump will run with full speed.
Shenchen Pump N6 Series 4. N6 Series Operation Interface 4.1 Main Interface 4.2 Menu Setting Interface - 3 -...
Page 7
Shenchen Pump N6 Series Steps: A. Press [digital knob] in the main interface to enter the menu setting interface, press again to select the setting parameters. B. Turn the [digital knob] to move up and down to select the parameter to be set.
Page 8
Shenchen Pump N6 Series F. Reset function: Select ‘Yes’ and press[ digital knob], the ‘Reset calibration finished!’ dialog box will pop up, the calibrate factor been restored. 4.2.2 Select Pump Head & Tube Interface Steps: a. Select the pump head model by pressing the [ digital knob] in Head & Tube interface.
Page 9
Shenchen Pump N6 Series Steps: A. Press [digital knob] in the dispensing interface, the interface turn to (1) state. Turn the knob to select setting On/Off, Time or Unit. B. In (1) state, press the knob change the interface to (2) state.
Page 10
Shenchen Pump N6 Series E. In the (2) state, press the [digital knob] or [ESC] to return to the (1) state and confirm the selected item. F. In the (1) state, turn the [ digital knob] clockwise to enter the external control speed customization interface, as shown below: G.
Page 11
Shenchen Pump N6 Series set 0V to 0 rpm, 5V to 600 rpm (2.5V should be 300 rpm), the maximum speed is set to 300 rpm, if the external input analog signal is 2.5V, then the motor speed is 300 rpm, input signal The motor speed remains unchanged at 300 rpm after exceeding 2.5V.
Page 12
F. In the (2) state, press the [ESC] or the [ digital knob] to return to the (1) state and confirm the setting parameters. Note: 1. N6 series pump support Modbus communication protocol-- RTH mode, communication port RS232 or RS485, baud rate 9600.
Page 13
Shenchen Pump N6 Series 4.2.7 Back Suction Interface Operation steps: A. In the (1) state, turn the [ digital knob] to select setting on/off or suction angle. B. In the (1) state, press the [ digital knob] change to the (2) state.
Shenchen Pump N6 Series 6. N6 Series Main Functions & Features 3.2 inch big screen LCD display working parameters in real time Super silent drive setting, precise control, low vibration and low low noise. Mechanical keypad and digital knob control, convenient for users setting ...
Shenchen Pump N6 Series N6-12L/YZ35 8. External Control Instruction - 14 -...
Page 18
Shenchen Pump N6 Series ① Analog signal input port: Turn on external control speed function in external control setting interface, control the pump speed from 0rpm to maximum speed through analog signal. 0-10V: 0V to 10V voltage signal input terminal.
Page 19
Shenchen Pump N6 Series External control wiring connection as below: A. In pulse mode: Disconnect K2 after short connect, the motor start running, disconnect again after short connect, motor stop running. In level mode: Short connect K2, motor start running; disconnect K2, motor stop running.
Page 20
Shenchen Pump N6 Series to the state before running at full speed. ( External control full speed default is level mode. ) ⑤ The motor working status output terminal: Output motor working status as below: If connect with external relay, when motor is running, K1 is connect;...
Page 21
Shenchen Pump N6 Series ⑦ RS485 communication terminal: Select RS485 in communication setting interface, then this terminal is effective. GD1: RS485 signal ground A+: Connect with RS485 A+ terminal B-: Connect with RS485 B- terminal - 18 -...
Shenchen Pump N6 Series 9. Maintenance When the pump is not working, loosen the tubing, avoid the tubing get plastic deformation because of long time pressure. The pump drive and pump head do not resist strong corrosive liquids, please pay attention to this when use the pump.
Shenchen Pump N6 Series 10. Warranty and After Sales Service We support three years warranty (do not include tubing). During the warranty, the pump is damaged by wrong operation or man-made sabotage, our company do not responsible for warranty. If the product out of warranty, we only charge cost for the replacement parts and shipping cost.
Page 24
MADE IN CHINA Baoding Shenchen Precision Pump Co., Ltd. Address:No.103,Building 2, ZhiDian Industrial Park, FuXing East Road 999, Baoding,China. Tel: 0086- 0312- 5958380 Fax: 0086-312- 6780636 Website: www.good-pump.com Email: info@good-pump.com...