Signal I/O
Input Signal
The following signals can be input when the ZFX-C is in the RUN mode.
Measurement Timing Input
A one-time measurement is performed if the TRIG signal is turned ON.
Command Input
Input format
DI8
DI7
Commands
Parallel command confirmation signal
Commands
Signals DI0 to DI7 can input the following commands. Allow at least 5 ms after DI0 to DI7 are determined to be
ON before turning DI8 ON.
Note
With the model re-registration setting commands, item Nos.0 to 31 can be registered. Item No.32 onwards cannot
be registered by these commands. They can be registered by serial communication commands.
Command
Continuous
measurement
Bank switching
Bank group
switching
Model
re-registration
Clear
measured data
162
Signal I/O
DI6
DI5
DI4
DI3
Description
Measurements continue while the com-
mand is being input.
Switches the bank to measure.
Switches the bank group data.
Registers the model again based on the
last measurement image that was read.
Clears the following measurement results:
• Judgment results and measurement values
of measurement items
• Judgment results and measurement values
of expressions
• Logging data
• Parallel output signals
(OR, DO0 to DO15)
DI2
DI1
DI0
Parameters
Command code Parameter (binary code)
DI7
DI6
0
0
0
1
1
1
1
0
1
0
DI5
DI4
DI3
DI2
DI1
0
*
*
*
*
0
Bank number
0
Bank group number
0
Item number
1
0
0
0
0
ZFX-C User's Manual
DI0
*
0