Technical specifications

12.1

Power

General
Supply voltage
Signal range
Startup current
Ordinary Location, Nonincendive, Dust Ignition Proof, Dust Protection by Enclosure Ex 't'
Voltage
Current
Instrinsically Safe Ex 'i'
U
/Vmax
i
I
/Imax
i
P
i
C
i
L
i
12.2

Temperature

Intrinsically safe
For gas application under the intrinsically safe protection Ex ia (EPL Ga / Category 1G / Class I, Div.
1), below is the ambient temperature de-rating table for various T codes and elevated process
temperatures.
T-Code
T6
T4, T5
SITRANS LR500 series with mA/HART
Operating Instructions, 03/2024, A5E51099898-AA
Ambient Temperature
-40 to +65 °C (-40 to +149 °F)
-40 to +45 °C (-40 to +113 °F)
-40 to +40 °C (-40 to +104 °F)
-40 to +80 °C (-40 to +176 °F)
-40 to +65 °C (-40 to +149 °F)
-40 to +55 °C (-40 to +131 °F)
30 V DC
4 to 20 mA
3.6 mA
24 V DC Nom, 30 V DC Max
4 to 20 mA
30 V
120 mA
0.8 W
2.2 nF
0
Process Temperature
-40 to +65 °C (-40 to +149 °F)
-40 to +160 °C (-40 to +320 °F)
-40 to +250 °C (-40 to +482 °F)
-40 to +80 °C (-40 to +176 °F)
-40 to +160 °C (-40 to +320 °F)
-40 to +250 °C (-40 to +482 °F)
12