Parameter assignment
9.2 Setup (2)
Note
"Upper range value" (URV) versus "Upper scaling point"
• When parameter "Upper range value" is set within the "Quick commissioning wizard", it
automatically sets the "Upper scaling point" to the same value.
• Setting a value for either parameter outside of the wizard, will not automatically adjust the
other value.
9.2.6

Custom (2.6)

Note
Menu visibility
This menu will not appear on device unless it is configured.
9.2.6.1

Upper scaling point (2.6.1)

Sets maximum scaled measurement value.
Setting
Default
Note
"Upper range value" (URV) versus "Upper scaling point"
• When parameter "Upper range value" is set within the "Quick commissioning wizard", it
automatically sets the "Upper scaling point" to the same value.
• Setting a value for either parameter outside of the wizard, will not automatically adjust the
other value.
9.2.6.2

Customized characteristic curve (2.6.2)

Used to enter Level and Output breakpoints.
If the shape of vessel (volume) is more complex than any of the preconfigured shapes, you can
define the shape as a series of segments. A value is assigned to each input (level) breakpoint and
a corresponding value is assigned to each output (volume) breakpoint.
• Level values are defined in Units (2.2.1) (Page 116).
• Volume values are defined in Volume units (2.5.4) (Page 125).
For more information, see Volume units (2.5.4) (Page 125).
126
0 to 9999999
100 <Custom units>
SITRANS LR500 series with mA/HART
Operating Instructions, 03/2024, A5E51099898-AA

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sitrans lr510Sitrans lr550Sitrans lr530Sitrans lr580

Table of Contents