Siemens SITRANS LR500 Series Operating Instructions Manual
Siemens SITRANS LR500 Series Operating Instructions Manual

Siemens SITRANS LR500 Series Operating Instructions Manual

Radar level transmitters
Hide thumbs Also See for SITRANS LR500 Series:
Table of Contents

Advertisement

Quick Links

SITRANS L
Radar level transmitters
SITRANS LR500 series with mA/
HART
Operating Instructions
7ML571..(LR500 Threaded lens antenna) 7ML573..
(LR500 Flanged encapsulated antenna) 7ML575..
(Polymeric horn antenna) 7ML578..(Flanged lens
antenna)
03/2024
A5E51099898-AA
Getting started
Introduction
Safety notes
Description
Installing/mounting
Connecting
Commissioning
Operating
Parameter assignment
Service and maintenance
Diagnostics and
troubleshooting
Technical specifications
Dimension drawings
Product documentation
and support
Technical reference
Communication
Remote operation
1
2
3
4
5
6
7
8
9
10
11
12
13
A
B
C
D

Advertisement

Table of Contents
loading

Summary of Contents for Siemens SITRANS LR500 Series

  • Page 1 Getting started Introduction Safety notes SITRANS L Description Radar level transmitters SITRANS LR500 series with mA/ Installing/mounting HART Connecting Operating Instructions Commissioning Operating Parameter assignment Service and maintenance Diagnostics and troubleshooting Technical specifications Dimension drawings Product documentation and support 7ML571..(LR500 Threaded lens antenna) 7ML573..
  • Page 2 Note the following: WARNING Siemens products may only be used for the applications described in the catalog and in the relevant technical documentation. If products and components from other manufacturers are used, these must be recommended or approved by Siemens. Proper transport, storage, installation, assembly, commissioning, operation and maintenance are required to ensure that the products operate safely and without any problems.
  • Page 3: Table Of Contents

    Local governing regulations ....................22 Use in hazardous areas....................... 23 Radar frequencies for worldwide use .................. 24 FCC Conformity........................25 Information for radio approval FCC and ISED............... 25 Radio equipment compliance (Europe) ................29 SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 4 Connecting or disconnecting in explosive environments ............. 50 6.1.3 Note on electromagnetic compatibility ................51 6.1.4 Improvement of interference immunity ................52 Connecting SITRANS LR500 series ..................52 6.2.1 Wiring instructions......................52 6.2.2 Input supply cable note ...................... 53 6.2.3 Connecting field device and SITRANS AW050 Bluetooth adapter..........
  • Page 5 General requirements ......................62 Local operation........................62 Device startup ........................63 Gather parameter values before running quick start wizards ..........64 Local commissioning......................64 7.6.1 Activating SITRANS LR500 series..................64 7.6.2 Local HMI .......................... 65 7.6.2.1 The LCD display........................66 7.6.3 Programming........................
  • Page 6 Start view (2.7.1)......................127 9.2.7.2 Contrast (2.7.2) ....................... 128 9.2.7.3 Display test (2.7.3)......................128 Maintenance and diagnostics (3)..................128 9.3.1 Identification (3.1) ......................129 9.3.1.1 Tag (3.1.1)........................129 9.3.1.2 Long tag (3.1.2)....................... 129 SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 7 Loop test (3.8.1) ......................143 9.3.8.2 Terminal voltage (3.8.2)....................143 9.3.9 Resets (3.9) ........................143 9.3.9.1 Device restart (3.9.1) ....................... 143 9.3.9.2 Resets (3.9.2) ........................144 9.3.9.3 Reset peak values (3.9.3) ....................144 SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 8 Common problems ......................165 11.6.1.1 Communication troubleshooting..................165 11.6.2 Measurement difficulties ....................166 11.6.2.1 Adjusting aiming ......................166 11.6.2.2 Fixed reading ........................166 11.6.2.3 Set the device to ignore the false echo ................167 SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 9 Echo selection ......................... 202 B.2.2.1 Confidence ........................202 B.2.2.2 Auto false echo suppression and Custom TVT ..............202 B.2.2.3 Echo selection algorithms ....................202 Loop power ........................203 B.3.1 Loop power ........................203 SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 10 Information for radio approvals ..................214 HART ..........................215 D.3.1 HART communication ...................... 215 D.3.2 Communication (HART) ....................215 D.3.3 Communication connections.................... 216 D.3.4 Configuring communication ports ..................217 D.3.5 Communication troubleshooting..................217 Index ..............................219 SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 11: Getting Started

    Installing/mounting (Page 35) 2. Connect the device. Connecting (Page 49) 3. Power up the device. Activating SITRANS LR500 series (Page 64) 4. Configure the device via quick commissioning wizard: Quick commissioning (Page 69) You can find additional functions in the section Parameter assignment. (Page 113)
  • Page 12 Getting started 1.1 Getting started SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 13: Introduction

    Compatible version of device integration package edition 03/2024 First edition HART SIMATIC PDM V9.2 EDD: 1.00.00 or later   FW: 1.00.00 or later HW: 1.00.00 or later Device revision 1 or later SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 14: Designated Use

    WARNING Using a damaged or incomplete device Risk of explosion in hazardous areas. • Do not use damaged or incomplete devices. Items supplied • SITRANS LR500 series radar LR510 threaded lens   level transmitter antenna    ...
  • Page 15: Industrial Use Note

    Siemens’ products and solutions undergo continuous development to make them more secure. Siemens strongly recommends that product updates are applied as soon as they are available and that the latest product versions are used. Use of product versions that are no longer supported, and failure to apply the latest updates may increase customer’s exposure...
  • Page 16: Security Note

    Introduction 2.12 Notes on warranty To stay informed about product updates, subscribe to the Siemens Industrial Security RSS Feed under https://www.siemens.com/cert (https://www.siemens.com/cert). 2.10 Security note NOTICE Unauthorized product information or software Use only authorized Siemens websites when accessing any product information or software, including firmware updates, device integration files (EDD, for example), as well as other product documentation.
  • Page 17 Introduction 2.12 Notes on warranty The content reflects the technical status at the time of publishing. Siemens reserves the right to make technical changes in the course of further development. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 18 Introduction 2.12 Notes on warranty SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 19: Safety Notes

    • National Electrical Code (NEC - NFPA 70) (USA) • Canadian Electrical Code (CEC Part I) (Canada) Further provisions for hazardous area applications are for example: • IEC 60079-14 (international) • EN 60079-14 (EU and UK) SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 20: Conformity With European Directives

    • Only carry out modifications that are described in the instructions for the device. Failure to observe this requirement cancels the manufacturer's warranty and the product approvals. Do not operate the device after unauthorized modifications. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 21: Industry Canada

    – Devices shall be installed only at fixed locations. Devices shall not operate while being moved or while inside a moving container. – Hand-held applications and residential use are prohibited. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 22: Requirements For Special Applications

    Note Operation under special ambient conditions We highly recommend that you contact your Siemens representative or our application department before you operate the device under special ambient conditions as can be encountered in nuclear power plants or when the device is used for research and development purposes.
  • Page 23: Use In Hazardous Areas

    • Connect the device with type of protection "Intrinsic safety" solely to an intrinsically safe circuit. • Observe the specifications for the electrical data on the certificate and/or in Technical specifications (Page 169). WARNING Substitution of components Substitution of components may impair Intrinsic Safety. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 24: Radar Frequencies For Worldwide Use

    Selecting the frequency for your country group Operating the device without selecting the frequency for the appropriate country group constitutes a violation of the regulations of the radio approvals of the respective country. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 25: Fcc Conformity

    L’exploitation est autorisée aux deux conditions suivantes: (1) l’appareil ne doit pas produire de brouillage; (2) l’utilisateur de l’appareil doit accepter tout brouillage radioélectrique subi, même si le brouillage est susceptible d’en compromettre le fonctionnement. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 26 • The use of this device is on a "no-interference, no-protection" basis. That is, the user shall accept operations of high-powered radar in the same frequency band which may interfere with or damage this device SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 27 • Ce dispositif ne peut être exploité qu'en régime de non-brouillage et de non-protection, c'est-à-dire que l'utilisateur doit accepter que des radars de haute puissance de la même bande de fréquences puissent brouiller ce dispositif ou même l'endommager SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 28 FCC 15.256 statement This device is approved for unrestricted use only inside closed, stationary vessels made of metal, reinforced fiberglass or concrete. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 29: Radio Equipment Compliance (Europe)

    20 cm from all persons. Radio equipment compliance (Europe) Hereby, Siemens declares that the SITRANS LR500 series is in compliance with the essential requirements and other relevant provisions of Directive 2014/53/EC. The declaration of conformity may be consulted here (http://www.craf.eu/radio-observatories-in-europe/).
  • Page 30 • This device must be installed to provide a separation distance of at least 20 cm from people to ensure compliance with the RF exposure requirements. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 31: Description

    Description Description SITRANS LR500 series radar level transmitter with HART, 4 to 20 mA is ideal for level and volume measurements. It is suitable for liquids, slurries, and bulk solids. LR510 threaded lens antenna LR530 flanged encapsulated antenna   LR580 flanged lens antenna LR550 polymeric horn antenna  ...
  • Page 32: System Configuration

    System configuration ① LR510 ② power supply ③ ④ R=250 Ohms ⑤ HART modem ⑥ HART communicator Communication This device supports various communication protocols. For more information, see Communication (Page 205). SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 33 Description 4.4 Communication Hart communication Use shielded cables when HART communications applications are in a harsh EMC environment. For more information, see HART communications (Page 205) SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 34 Description 4.4 Communication SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 35: Installing/Mounting

    Note Pressure Equipment Directive (PED) 2014/68/EU Siemens Level Transmitters with flanged, threaded, or sanitary clamp type process mounts have no pressure-bearing housing of their own. Therefore, they do not come under the Pressure Equipment Directive as pressure or safety accessories (see EU Commission Guideline A-08 and A-20).
  • Page 36 Note Material compatibility Siemens can provide you with support concerning selection of parts wetted by process media. However, you are responsible for the selection of parts. Siemens accepts no liability for faults or failures resulting from incompatible materials. WARNING Unsuitable connecting parts Risk of injury or poisoning.
  • Page 37: Installation Location Requirements

    • Observe the maximum permissible ambient temperature. Refer to the information in the section Technical specifications (Page 169). NOTICE Aggressive atmospheres Damage to device through penetration of aggressive vapors. • Ensure that the device is suitable for the application. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 38: Direct Sunlight

    • Before installing ensure there is no visible damage to the device. • Make sure that process connectors are clean, and suitable gaskets and glands are used. • Mount the device using suitable tools. Refer to the information in Technical specifications (Page 169). SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 39: Incorrect Disassembly

    PTFE gasket is part of the antenna system, and which are suitable for the service conditions. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 40: Sensor Reference Point

    Ensure when planning your measuring point that the radar sensor has a " clear view" to the measured product. In case of existing vessel installations, an auto false echo suppression should be carried out during setup when the vessel is empty. > 200 mm (7.87") SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 41 G1, 1 NPT 10° G1½, 1½ NPT (250 °C) 10° G1½, 1½ NPT (150/200 °C) 7° LR550 DN 80, 3" 3° DN 100, 4" DN 150, 6" DN 200, 8" DN 250 SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 42: Nozzle Mounting

    LR510 threaded connection Nozzle diameter "A" Nozzle height "B" 40 mm 1 1/2" ≤ 150 m ≤ 5.9" 50 mm 2" ≤ 200 mm ≤ 7.9" 80 mm 3" ≤ 300 mm ≤ 11.8" SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 43: Thread Sealing

    (The maximum recommended torque for Threaded versions is 40 N-m (30 ft.lbs.) Note G thread types G thread types are supplied with a Klingersil flat seal. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 44: Lr550 Polymeric Horn, Liquids, And Solids

    5.5 LR550 polymeric horn, liquids, and solids LR550 polymeric horn, liquids, and solids LR550 design LR550 with adapter flange LR550 with DN80/3" Universal flange LR550 purging connection with adaptor flange, mounted on a nozzle SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 45: Mounting Bracket

    The optional mounting bracket allows simple mounting of the LR550 on a wall or ceiling. Especially in the case of open vessels, this is a simple and effective way to align the sensor to the surface of the bulk solid material. Ceiling mounting Wall mounting SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 46 Required tools: Allen wrench size 4. There are two different ways of screwing the bracket to the sensor, see following illustration: ① For angle of inclination, in steps ② For angle of inclination, infinitely variable Mounting bracket aiming SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 47: Sunshield

    Installing/mounting 5.6 Sunshield Sunshield There is an optional sunshield to protect the display, if the device is mounted in direct sunlight. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 48 Installing/mounting 5.6 Sunshield SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 49: Connecting

    Risk of explosion in hazardous area Depending on the device version, connect the power supply as follows: • Connecting terminals: Connect the terminals according to the terminal connection diagram. First connect the PE/ground conductor. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 50: Connecting Or Disconnecting In Explosive Environments

    • Connect or disconnect devices in hazardous areas only in a de-energized state. • Install a suitable switch-off device. Exceptions: • Devices having the type of protection "Intrinsic safety Ex i" may also be connected in energized state in hazardous areas. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 51: Note On Electromagnetic Compatibility

    Electromagnetic compatibility (EMC) You can use this device in commercial, industrial or business environments. Metal enclosures ensure improved protection from electromagnetic radiation. This protection can be increased by grounding the enclosure. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 52: Improvement Of Interference Immunity

    1. To remove lid from the device, turn by hand in a counter-clockwise direction. 2. Strip cable jacket for approximately 70 mm (2.75 inch) from end of cable, and thread wires through gland. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 53: Input Supply Cable Note

    Housing can be rotated beyond 360° without damaging the device. 6.2.2 Input supply cable note Note Insulation thickness The input supply cable should have an insulation thickness of at least 0.5 mm. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 54: Connecting Field Device And Sitrans Aw050 Bluetooth Adapter

    The cables are best run separately inside the enclosure as shown in the figure following. Take care when installing the display to be sure that the cables are not pinched or torn. Excessive force should not be necessary for installation of the display mounting. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 55: Installing Or Replacing Bluetooth Adapter

    3. Slide the adapter into the gland against the O-ring. 4. Tighten the gland to the adapter, without twisting the cables. 5. Plug the cables into the display and electronics. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 56: Hazardous Area Installations

    Note Sample nameplate This nameplate is given as an example only. SITRANS LR510 HART 7ML7510-0BC00-4AF3 The ATEX certificate number (CSANe 23ATEX1113X) listed on the nameplate can be downloaded from our website: SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 57: Dust Ignition

    Product page (www.siemens.com/sitransLR500) Go to Technical Info > Images, graphics, drawings. 6.3.1.2 Dust ignition Device nameplate DI Note Sample nameplate This nameplate is given as an example only. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 58 Product page (www.siemens.com/sitransLR500) The CSA certificate (80144219) listed on the nameplate can be downloaded from our website: Product page (www.siemens.com/sitransLR500) For more information on hazardous area installations, refer to Hazardous area installations (Page 56). SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 59: Non-Incendive

    • The supply to the SITRANS LR510, LR530, LR550, LR580 equipment shall be rated for a prospective short-circuit current of not more than 1.5 kA and shall be protected by a suitably rated fuse. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 60: Instructions Specific To Hazardous Area Installations (Reference Atex Directive 2014/34/Eu, Annex Ii, 1.0.6 And Uk Regulations Si 2016/1107))

    – Aggressive substances: for example, acidic liquids or gases that may attack metals, or solvents that may affect polymeric materials. – Suitable precautions: for example, establishing from the material’s data sheet that it is resistant to specific chemicals. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 61: Commissioning

    The same risk continues to apply when error messages are switched off or disabled. WARNING Hot surfaces Risk of burns resulting from hot surfaces. • Take corresponding protective measures, for example by wearing protective gloves. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 62: General Requirements

    • To have device process values update when configured remotely, use local buttons to take device in and out of measurement view (i.e. go to parameter view then back to measurement view). SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 63: Device Startup

    The next HMI view will be the operation view 1. For any subsequent startup, after power-on, the device automatically starts in operation view. A transition screen showing first the Siemens logo and then the current firmware revision of the product is displayed while the first measurement is being processed.
  • Page 64: Gather Parameter Values Before Running Quick Start Wizards

    Local commissioning SITRANS LR500 series can be quickly commissioned using wizards, and menu driven parameters. The parameters can be modified locally using the device display and buttons, also known as the Human Machine Interface (HMI).
  • Page 65: Local Hmi

    Note Backlight is always on The SITRANS LR500 series has a backlight which is always on to assist viewing in very low light conditions. This backlight switches off if the voltage is less than 14v. Note Show menu numbers on local display To view menu numbers on the device, parameter "Service view"...
  • Page 66: The Lcd Display

    A band halfway down the menu bar indicates the current item is halfway down the list. ① ④ Current menu Menu bar ② ⑤ Current item number Item band ③ Current item     SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 67: Programming

    6. Press to accept the new value. The display returns to parameter view ("EDIT" symbol is no longer flashing) and shows the new selection. Review for accuracy. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 68: Wizards

    In the upper left corner of each view, the wizard name appears, followed by each step name. In the upper right corner, the view number is shown. The number of steps depends on the actual device variant ordered. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 69: Quick Commissioning

    This device provides a single quick commissioning wizard that can be used for various applications. The initial wizard steps are common for all application types. Subsequent wizard parameters will vary depending on the application you choose. For the purpose of documenting, two SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 70 • To have device process values update when configured remotely, use local buttons to take device in and out of measurement view (i.e. go to parameter view then back to measurement view). SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 71 In the following wizard steps, a full list of options for each parameter is shown. However, some options may not appear on the device, based on the application selected. Operation Sets operation mode which defines output and local display. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 72 Lower calibration point (process empty level) Space   Upper calibration point (process full level) Distance to material surface Distance Sensor reference point   Custom Linearization table (level/volume Lower calibration point breakpoints) SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 73 However, if the value is too large to show on the segment display, the decimal places will be automatically adjusted to show the process value. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 74 Slow Damping value Used in the damping (filtering) of process values to smooth out response to sudden changes in measurement. Sets time after which output signal reaches 63% of final value. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 75 If a change is made to custom units, be sure to rescale the output, as rescaling for custom units is not handled automatically by the device. • To rescale output, use parameters "Upper range value" and "Upper scaling point". SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 76 In the following wizard steps, a full list of options for each parameter is shown. However, some options may not appear on the device, based on the application selected. Operation Sets operation mode which defines output and local display. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 77 Lower calibration point (based on level)   Material type Used to optimize performance based on material type. Setting Liquid Solid Default Liquid Application type Used to optimize performance based on application type. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 78 • Note that a change to custom units also requires a scaling adjustment (see Upper range value (Page 80)). Vessel shape Sets vessel shape, and allows device to calculate volume in addition to level. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 79 Sets distance from sensor reference point to upper calibration point: usually process full level. Response rate Sets reaction speed of device to measurement changes in target range. Use a setting just faster than the maximum filling or emptying rate (whichever is faster). SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 80 • When parameter "Upper range value" is set within the "Quick commissioning wizard", it automatically sets the "Upper scaling point" to the same value. • Setting a value for either parameter outside of the wizard, will not automatically adjust the other value. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 81: Demo Wizard

    Applies settings as last step in wizard. Setting Wizard completes and settings are applied. Returned to start of wizard. (Settings are saved to perform wizard again, but not applied until "Confirm" set to "Yes".) Default   SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 82: Afes Wizard

    Confirm - settings Setting Wizard completes and settings are applied. Returned to start of wizard. (Settings are saved to perform wizard again, but not applied until "Confirm" set to "Yes".) Default   SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 83: Requesting An Echo Profile

    Device address Setting a device address is not necessary for local operation, but must be set if configuring the device for use on a HART network. See Polling address (4.1) (Page 145). SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 84: Testing The Configuration

    The latter is preferred, as it more accurately represents operating conditions. However, if it is not possible to do a physical test, a simulation will ensure that control programming is correct. For more details, see Simulation (Page 89). SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 85: Operating

    Dimensions of the vessel, wet well, or reservoir (other than lower and upper calibration points) are only important if you require volume readings. In this case, all dimensions are used to calculate the volume value in terms of level. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 86: Ma Control

    Up to 32 breakpoints, each consisting of an input and output value (X- and Y-value), are available to define your vessel shape. • For a custom volume characteristic curve, X-values refer to level, Y-values refer to volume. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 87 ("Upper scaling point" is set by parameter "Upper range value" in "Quick commissioning wizard".) Example chart for custom volume characterization (with 15 of possible 32 breakpoints defined): SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 88 Y-value 11 level breakpoints. Y-value 12 Y-value 13 10.9 Y-value 14 13.0 Y-value 15 15.0 For more details regarding characterization, refer to Vessel shape and dimensions (Page 89). SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 89: Vessel Shape And Dimensions

    HMI, or via an engineering system such as SIMATIC PDM. Process values and diagnostics can be simulated via a remote engineering system. Current output simulation The current output can also be simulated using Loop test (3.8.1) (Page 143). SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 90: Simulation - Process Value (Operating Chapter)

    Simulation - Process value (Operating chapter) Process value simulation Process value simulation is an iterative process whereby parameters are adjusted and corresponding results are displayed. Process values can be simulated using a remote SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 91 1. Set parameter "Simulation mode" to "Ramp" to simulate a changing process value. 2. Set parameter "Simulation value" to the desired starting value for the simulation. 3. Set parameter "PV status" to the status to simulate. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 92: Simulation - Application Test (Operating Chapter)

    Diagnostics can be simulated via a remote engineering system. Diagnostics related to "Device status", "Limit monitoring and event counter", and "HART status" can be chosen from a select list to simulate. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 93: Remote Operation

    SIMATIC PDM Version Check the support page of our website to make sure you have the latest version of SIMATIC PDM, the most recent Service Pack (SP) and the most recent hot fix (HF): Siemens Industry Online Support (https://support.industry.siemens.com/cs/?lc=en-WW) Browse to Automation Technology >...
  • Page 94: Updating The Electronic Device Description (Edd)

    LR500. The EDD revision must match the Firmware revision in the device. Note SITRANS LR500 series standard PDM EDD and Field Device Integration (FDI) EDD FDI EDD may have limited functionality in comparison to the standard PDM EDD. To install a new EDD: 1.
  • Page 95: Quick Start Note

    • For a vessel with obstructions see Auto False Echo Suppression via SIMATIC PDM (Page 98). 8.6.1.7 Quick start menu and instruction Launch SIMATIC PDM, open the menu Device – Wizard - Quick Start, and follow steps 1 to 4. ① Quick start wizard SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 96 Choose a response rate setting which is slightly faster than the actual process rate. This setting will affect other parameters such as the fill and empty rate, damping and window discrimination method. Step 2 – Application SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 97: Changing Parameter Settings Using Simatic Pdm

    2. Adjust parameter values in the parameter value field then Enter. The status fields read Changed. 3. Open the Device menu, click on Download to device, then use File - Save to save settings offline. The status fields are cleared. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 98: Parameters Accessed Via Pull-Down Menu

    The device learns the echo profile over the whole measurement range and the TVT is shaped around all echoes present at that moment. See Auto False Echo Suppression (Page 202) for a more detailed explanation. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 99 8. To turn Auto False Echo Suppression off or on, reopen the Auto False Echo Suppression window, change the Auto False Echo Suppression to Off or On, click on Transfer to Device. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 100 ③ False echo     After Auto False Echo Suppression ① ④ False echo Material level ② ⑤ Learned TVT Default TVT ③ ⑥ Auto False Echo Suppression Range Echo marker SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 101 If the TVT Shaper tab is closed without transferring new values to the device, the new values will appear in the PDM list even though they have not been transferred to the device. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 102 Click on OK to proceed. The new profiles will be saved with their date and time. • Click on the tab View Saved Echo Profiles to view the stored profiles. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 103 Lifetime (Required)/Reminder 2 before Lifetime (Demanded). 3. Click Write. 4. Click Read, to see the effects of your modification. 5. Click Snooze to add a year to the Total Expected Device Life. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 104 3. When you are ready to end simulation, select End and click on OK to return the device to the actual output value. Configuration flag reset To reset the configuration flag to zero, open the menu Device – Configuration Flag Reset and perform a reset. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 105 Overview to see reading (level, space, distance); analog output; device status; and current electronics temperature. To see highest and lowest electronics temperatures, navigate to Level Meter > Maintenance and Diagnostics > Electronics Temperature. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 106 Open the menu View – Process Variables and click on Trend. The mA Output follows Primary Value (PV), and can be set to either Level, Space, or Distance. PV selection (2.1.1) (Page 114) SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 107 With the limit monitoring and event counter function, the following options are available to you via remote operation (e.g. SIMATIC PDM): • Monitoring process values • Counting events based on configured limits • Triggering, acknowledging and resetting process value alarms and warnings. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 108 If the monitored process value is again within the limit values, the process value alarm is reset. Configuring the event counter Condition You have configured the following values in limit monitoring: • Upper limit • Lower limit • Hysteresis Configuring limit monitoring (Page 108) SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 109 1. Select the menu command "Device > Limit monitoring and event counter". 2. Click "Reset and acknowledge". Result Process value alarms and warnings are acknowledged and deleted. The event counter is reset. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 110 The next underrun event is counted when the process value first rises above the lower limit ② ③ plus the entered hysteresis , and then falls below the lower limit Within 'B', two events are counted in the displayed period. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 111 5. Select the process values you want to record. 6. Click on "Transfer" to write the log settings to the device. The buffer with the existing logging points is deleted and overwritten with new logging points. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 112 1. Use remote operation (e.g. SIMATIC PDM) to select the menu command "Diagnostics > Device status". 2. Select the "HART status" tab. The operating time and, if available, the sensor operating time is displayed. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 113: Parameter Assignment

    • See Quick commissioning via SIMATIC PDM (Page 207). Note No user-relevant data is lost in the event of a power failure. 9.1.1 Quick commissioning (1.1) Provides step-by-step procedure to set up common applications easily. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 114: Demo Wizard (1.2)

    • Options "Volume" and "Custom" are not visible in measurement view until configured. If one of these unconfigured options is selected in parameter "Start view", measurement view will show the next visible process value. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 115: Selection (2.1.2)

      Space Distance Signal strength Confidence Volume Custom Electronics temperature Default Distance 9.2.1.3 TV selection (2.1.3) Sets the tertiary variable by selecting a process value that corresponds to the loop current. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 116: Qv Selection (2.1.4)

    Electronics temperature 9.2.1.5 Linearization type (2.1.5) Sets type of linearization used to calculate volume. Setting None Volume Custom Default None 9.2.2 Sensor (2.2) 9.2.2.1 Units (2.2.1) Sets units used by the device. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 117: Temperature Units (2.2.2)

    °C (degrees Celcius) • K (Kelvin) • °F (degrees Fahrenheit) • °R (degrees Rankine) Default °C (degrees Celcius) 9.2.2.3 Fill rate limit (2.2.3) Sets the maximum fill rate the device can track. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 118: Empty Rate Limit (2.2.4)

    0 to 120 m Default 30 m 9.2.3.2 Upper calibration point (2.3.2) Sets distance from sensor reference point to upper calibration point: usually process full level. Setting 0 to 120 m Default SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 119: Lower Level Point (2.3.3)

    By using the default value of -999, or a lower value, the parameter "Low-level cutoff" is disabled. 9.2.3.7 Propagation factor (2.3.7) Sets value that represents reduction in propagation velocity as a result of the wave travelling through a pipe. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 120: Antenna Offset (2.3.8)

    Parameter assignment 9.2 Setup (2) Setting 0.3 to 1.1 Default 9.2.3.8 Antenna offset (2.3.8) Sets offset required if electronics module is replaced. Consult your local Siemens representative for support. Setting -0.2 to 0.2 m Default 0.0 m 9.2.4 Current output (2.4) 9.2.4.1...
  • Page 121: Lower Range Value (2.4.4)

    • PV set to Volume - Setting based on parameter "Upper scaling point" in "Volume" menu • PV set to Custom - Setting based on parameter "Upper scaling point" in "Custom" menu SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 122: Lower Saturation Limit (2.4.6)

    Setting 20 to 22.8 mA Default 20.5 mA 9.2.4.8 Lower fault current (2.4.8) Sets Lower fault current in Non-safety mode. Setting 3.55 to 4.0 mA Default 3.55 mA SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 123: Upper Fault Current (2.4.9)

    Note Menu visibility This menu will not appear on device unless it is configured. 9.2.5.1 Vessel shape (2.5.1) Sets vessel shape, and allows device to calculate volume in addition to level. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 124 Horizontal cylinder flat ends vessel Horiz.cyl.parab.ends Upper scaling point, Vessel dimension A, Vessel dimension L Horizontal cylinder parabolic ends vessel Sphere Upper scaling point Sphere vessel Default Linear ① ② Dimension A Dimension L SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 125: Vessel Dimension A (2.5.2)

    (US) Normal liters Normal cubic meters Standard cubic feet Default Cubic meters 9.2.5.5 Upper scaling point (2.5.5) Sets maximum scaled measurement value. Setting 0 to 9999.999 m Default 1000.0 m SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 126: Custom (2.6)

    (volume) breakpoint. • Level values are defined in Units (2.2.1) (Page 116). • Volume values are defined in Volume units (2.5.4) (Page 125). For more information, see Volume units (2.5.4) (Page 125). SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 127: Local Display (2.7)

    9.2.7.1 Start view (2.7.1) Sets process value shown first on display after power on. Setting Level Space Distance Volume Custom Loop current Primary variable Percent of range Electronics temperature Default Level SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 128: Contrast (2.7.2)

    • 90 % • 100 % Default 50 % 9.2.7.3 Display test (2.7.3) Used to test device display. Maintenance and diagnostics (3) The following parameters relate to device maintenance and diagnostics. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 129: Identification (3.1)

    Displays the article number (MLFB) of the device. Serial number (3.1.5.4) Displays unique serial number of device, set by factory. HW version (3.1.5.5) Displays version number corresponding to electronics hardware of device. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 130: Local Display (3.1.6)

    [ ] Sensor break [ ] Out of service [ ] Invalid device configuration [ ] Sensor fail-safe timer expired Default [ ] indicates active device status [ ] indicates inactive device status SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 131: Device Status 2 (3.2.3)

    [ ] PV status: uncertain [ ] PV status: bad [ ] Device lifetime: maintenance required [ ] Device lifetime: maintenance demended Default [ ] Indicates active device status [ ] Indicates inactive device status SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 132: Device Status 4 (3.2.5)

    Signal quality (3.3.1) Confidence (3.3.1.1) Displays echo quality: higher values represent better echo quality. Echo signal strength (3.3.1.2) Displays strength of selected echo, in dB. Noise average (3.3.1.3) Displays the average ambient noise. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 133: Echo Configuration (3.3.2)

    Sets minimum echo confidence. Only echoes above this threshold are evaluated. Setting -20 to 70 Default 1 to 70 Position detect (3.3.3.3) Sets algorithm used to calculate the position of the echo. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 134: Filtering (3.3.4)

    0 to 255 Default Note If Near range suppression is enabled, Auto false echo suppression (Page 98) must be disabled. Near range suppression distance (3.3.4.3) Sets distance for near range suppression. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 135: Sampling (3.3.5)

    Sets measurement verification process. When echo lock set to "Total lock", parameter "Echo lock window" is preset to 0 for automatic width and cannot be changed. Setting Disabled Maximum verification Material agitator Total lock Default Material agitator SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 136: Tvt Configuration (3.3.6)

    Use "Material agitator" option to prevent false echoes from agitator blades. "Maximum verification" option can be used if experiencing frequent loss of echo (LOE). However, it is recommended that you contact your Siemens representative before activating. Echo lock window (3.3.5.2) Sets distance window (centered on echo) used to derive reading.
  • Page 137 Breakpoint 61 (3.3.6.8.1) ... Breakpoint 90 (3.3.6.8.30) Sets TVT offset for breakpoint on TVT. Shaping points can be raised (positive number) or lowered (negative number). Setting Offset -50 to 50 dB Default 0 dB SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 138: Peak Values (3.4)

    Minimum echo signal strength (3.4.5) Displays minimum echo signal strength value. 9.3.4.6 Maximum echo signal strength (3.4.6) Displays maximum echo signal strength value. 9.3.4.7 Minimum confidence (3.4.7) Displays minimum confidence value. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 139: Maximum Confidence (3.4.8)

    9.3.5.4 Maximum terminal voltage (3.5.4) Displays maximum terminal voltage value. 9.3.6 Trend log settings (3.6) 9.3.6.1 Number of logging values (3.6.1) Sets number of process values to log. Setting Disabled Default SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 140: Number Of Logged Points (3.6.2)

    Sets interval in seconds between log entries. Setting 10 to 600 Default 9.3.6.4 Logging behavior (3.6.4) Defines behavior when log is full. Setting Overwrite oldest Fill and stop Default Overwrite oldest SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 141: Logging Value 1 (3.6.5)

    To edit the logging values, trend logging must be disabled. 9.3.7 Simulation (3.7) 9.3.7.1 Simulation mode (3.7.1) Defines whether the simulated value is fixed or in the shape of a ramp. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 142: Simulation Value (3.7.2)

    Sets number of stair steps between ramp start (parameter "Simulation value") and end of ramp simulation. Setting 0 to 65535 Default 9.3.7.6 Ramp duration (3.7.6.) Sets total time from ramp start (parameter "Simulation value") to end of ramp simulation. Setting 0 to 65535 Default SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 143: Current Loop (3.8)

    Displays the voltage present at the device terminals. 9.3.9 Resets (3.9) 9.3.9.1 Device restart (3.9.1) Used to restart device without disconnecting power. Simulation will be terminated. Stored configurations are not reset. Setting Cancel Default Cancel SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 144: Resets (3.9.2)

    Resets all recorded Peak values. Setting Electronics temperature Distance Echo signal strength Confidence Peak values (Resets recorded minimum and maximum values for Distance, Electronics temper‐ ature, Echo signal strength, and Confidence.) Default SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 145: Frequency (3.10)

    The following parameters relate to device Communication. 9.4.1 Polling address (4.1) Sets poll ID on a HART network. For point-to-point configurations, the standard address is zero (0). For multidrop configurations, use a non-zero HART address. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 146: Identify The Device (4.2)

    Used to change PIN code that enables user access level. Setting 0 to 65535 Default 9.5.3 Recovery ID (5.2) Displays recovery ID that must be provided to Technical Support to obtain the PIN-Unlock-Key (PUK) required to recover the PIN(s). SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 147: Pin Recovery (5.3)

    EDD and saved in offline table, e.g. PDM Structure view.) When you provide this information to Siemens Technical Support, a PUK (PIN Unlock Key) will be supplied to you. Enter this PUK in parameter "PIN recovery" to reset the user PIN to factory default value.
  • Page 148: Button Lock (5.5)

    Disabled Language (6) Sets the language for the local display. Setting • English • Deutsch • Français • Italiano • Español • 汉语 Default Based on user configuration at initial startup. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 149: Service And Maintenance

    • Ensure that the atmosphere is explosion-free (hot work permit). NOTICE Penetration of moisture into the device Damage to device. • Make sure when carrying out cleaning and maintenance work that no moisture penetrates the inside of the device. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 150: Cleaning

    WARNING Impermissible repair of explosion protected devices Risk of explosion in hazardous areas • Repair must be carried out by Siemens authorized personnel only. WARNING Maintenance during continued operation in a hazardous area There is a risk of explosion when carrying out repairs and maintenance on the device in a hazardous area.
  • Page 151: Enclosure Open

    Risk of burns during maintenance work on parts having surface temperatures exceeding 70 °C (158 °F). • Take corresponding protective measures, for example, by wearing protective gloves. • After carrying out maintenance, remount touch protection measures. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 152: Replacing The Hmi

    Gently turn HMI one quarter turn clockwise to secure the HMI in the enclosure. 3. Replace device lid. Thread onto enclosure, turning clockwise. Hand tighten until mechanical stop is reached. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 153: Return Procedure

    To return a product to Siemens, see Return goods delivery note (https://www.siemens.com/ processinstrumentation/returngoodsnote). Contact your Siemens representative to clarify if a product is repairable, and how to return it. They can also help with quick repair processing, a repair cost estimate, or a repair report/ cause of failure report.
  • Page 154: Disposal

    Information about battery / product return (WEEE) (https:// support.industry.siemens.com/cs/document/109479891/) Note Special disposal required The device includes components that require special disposal. • Dispose of the device properly and environmentally through a local waste disposal contractor. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 155: Diagnostics And Troubleshooting

    – Local display (using Namur symbols) will show "Configuration error" as higher priority. – SIMATIC PDM (using Siemens standard symbols) will show "Maintenance demanded" as higher priority. Be aware of the priority for each device status, depending on the interface used.
  • Page 156: Device Status Symbols

    Button lock (5.5) (Page 148) is set to "Off", and that User PIN is deactivated. If you continue to experience problems, go to: Product page (www.siemens.com/LR500) Check the FAQs for this device, or contact your Siemens representative. 11.3 Device status symbols 11.3.1...
  • Page 157 Action: Check ambient temperature or process conditions. If possible, install device at different location.   Function check Configuration warning (yellow) Cause: Device can operate, but one or more parameters are incorrectly configured. Action:   Out of specification Process value warning   SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 158: Device Information Symbols

    Cause: Device status ok. No active diagnostic errors. Action: No action required. * Lowest priority number equals highest fault severity. ** Both the Siemens standard symbol and its corresponding Namur symbol (from device display) will be shown in SIMATIC PDM. 11.4 Device information symbols 11.4.1...
  • Page 159: Fault Codes And Corrective Actions

    The number of overruns of the process value (set in parameters "Upper limit" and "Monitored value") has reached the threshold. Number over‐ runs above Reset and acknowledge event counter. threshold Check process conditions. Maintenance Check limit monitoring and event counter settings. alarm SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 160 Forthcoming end of the calibration interval. maintenance re‐ Maintenance of device should be planned. quired Limit monitor‐ Monitored value is above limit (set in parameter "Upper limit"). ing 1   Above limit Process value alarm SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 161 "Monitored value") has reached the threshold. Number under‐ runs above Reset and acknowledge event counter. threshold Check process conditions. Process value Check limit monitoring and event counter settings. alarm SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 162 The number of overruns of the process value (set in parameters "Upper limit" and "Monitored value") has reached the threshold. Number over‐ runs above Reset and acknowledge event counter. threshold Check process conditions. Maintenance re‐ Check limit monitoring and event counter settings. quired SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 163 Make sure input voltage is within product specification. Maintenance demanded Supply voltage The supply voltage is too high. above limit Make sure input voltage is within product specification. Maintenance alarm SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 164 Repair is required. Contact Technical support. alarm   Volatile memo‐ Device electronics error. ry check failure Restart the device. Maintenance If error continues, device electronics may have a defect. alarm Repair is required. Contact Technical support. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 165: Operation Troubleshooting

    If you continue to experience problems, go to our Product page (www.siemens.com/LR500) and check the FAQs for SITRANS LR500, or contact your Siemens representative. SITRANS LR500 series with mA/HART...
  • Page 166: Measurement Difficulties

    If the value in parameter "Fail-safe LOE timer" expires due to a measurement difficulty, the value set in parameter "Fail-safe loss of echo" displays. In rare cases, the SITRANS LR500 series may lock on to a false echo and report a fixed or wrong reading.
  • Page 167: Set The Device To Ignore The False Echo

    “True first echo” algorithm. Also, if the true echo has jagged peaks, use parameter "Reform echo". Should a stable measurement still not be attainable, contact Siemens representative. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 168 Diagnostics and troubleshooting 11.6 Operation troubleshooting SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 169: Technical Specifications

    -40 to +65 °C (-40 to +149 °F) -40 to +160 °C (-40 to +320 °F) -40 to +55 °C (-40 to +131 °F) -40 to +250 °C (-40 to +482 °F) SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 170 -40 to +60 °C (-40 to +140 °F) -40 to +150 °C (-40 to +302 °F) -40 to +55 °C (-40 to +131 °F) -40 to +250 °C (-40 to +482 °F) SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 171: Performance

    +/- 10 mm (0.39 in) from reference point to 250 mm (9.84 in) distance From sensor reference point: see Dimensions (Page 187). Current output damping, number of shots averaging, fill & empty rate limit settings all reduce response time SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 172: Interface

    Standard, integral to analog output Configuration Local graphical display with backlight and pushbuttons. SITRANS mobile IQ App with AW050 module (non Ex devices only), Siemens SIMATIC PDM HMI (local) Graphical backlit liquid crystal, with bar graph (representing Level) Display will be degraded below -25 °C (-13 °F) and above 65 °C (149 °F) 12.5...
  • Page 173: Operating Conditions

    Suitable for outdoor (TYPE 4X, TYPE 6, IP66, IP68 enclosure) Installation category Pollution degree Storage and operation Storage period: 12 to 18 months (if product is kept clean, dry, and within temperature specifications)   Safe operation period: 15+ years SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 174: Process Conditions

    0.8 bar (11.6 psig) 3.3 m 1.5 m 1 bar (14.5 psig) 3.6 m 1.2 bar (17.4 psig) 3.9 m 2.3 m 1.4 bar (20.3 psig) 2.7 m 1.6 bar (23.2 psig) 4.3 m SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 175: Process

    -40 … +150 °C (-40 … +302 °F) Threaded connection -40 … +200 °C (-40 … +392 °F) FFKM -15 … +150 °C (5 … +302 °F) -15 … +250 °C (5 … +482 °F) SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 176: Certificates And Approvals

    • CSAUK 24JPN016X Ex ia IIIC T200 95°C Da • CSAUK 24JPN017X   Brazil     Ex ia IIC T6..T4 Ga • INMETRO DNV 24.0019X Ex ia IIIC T 95°C Da SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 177: Communication

    United Kingdom UKCA   LYH-LR500 •   267AA-LR500 • Canada     • Japan 12.10 Communication Communication type • HART 7 • 4 to 20 mA Supported engineering system SIMATIC PDM SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 178: Derating Curves

    Technical specifications 12.11 Derating curves 12.11 Derating curves 12.11.1 Temperature derating curves LR510 threaded lens antenna LR530 flanged encapsulated antenna SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 179 Technical specifications 12.11 Derating curves LR550 polymeric horn antenna LR580 flanged lens antenna SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 180: Temperature Vs Pressure Derating Curves

    LR510 threaded lens antenna Seal options 0 and 1 ① ④ Atmospheric pressure FKM without thermal extension ② ⑤ Allowable operating pressures FKM with thermal extension ③ Allowable operating temperatures     SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 181 Technical specifications 12.11 Derating curves Seal options 2 & 3 ① ④ Atmospheric pressure FKM without thermal extension ② ⑤ Allowable operating pressures FKM with thermal extension ③ Allowable operating temperatures SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 182 PTFE without thermal extension ② ⑤ Allowable operating pressures PTFE with thermal extension ③ Allowable operating temperatures     Note: Class 300 flange maximum allowable working pressure is 25 bar (362 psi). SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 183 Note: PN40 flange maximum allowable working pressure is 25 bar (362 psi). JIS B 2220 ① ④ Atmospheric pressure PTFE without thermal extension ② ⑤ Allowable operating pressures PTFE with thermail extension ③ Allowable operating temperatures     SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 184 LR580 flanged lens antenna Seal options 0 & 1 ① ④ Atmospheric pressure FKM without thermal extension ② ⑤ Allowable operating pressures FKM with thermal extension ③ Allowable operating temperatures     SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 185 Technical specifications 12.11 Derating curves Seal option 2 ① ④ Atmospheric pressure FFKM thermal extension ② Allowable operating pressures     ③ Allowable operating temperatures     SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 186 Technical specifications 12.11 Derating curves SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 187: Dimension Drawings

    Process seal Temperature ded max. material range m (ft) Thread G3/4" PN40, DIN3852-A 182.3 (7.18) 10 (32.8) 14° -40 ... +150 °C (-40 ... +302 Thread 3/4" NPT, ASME B1.20.1   °F) SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 188 Thread G1" PN40, DIN3852-2-A 245.8 (9.68) 20 (65.6) 10° Thread 1" NPT, ASME B1.20.1 246 (9.68) Thread G1-1/2" PN40, DIN3852-2-A 237.8 (9.36) 30 (98.4) 10° Thread 1-1/2" NPT, ASME B1.20.1 238 (9.37)     SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 189: Lr530 Flanged Encapsulated Antenna

    (8.73) (3.94) (0.43) (0.63) EN1092-1, DIN2501 / 316/316L flange DN50 PN6, raised 216.1 268.1 6° 30 (98.4) 110 (4.33) face, Form B1, (8.51) (10.56) (5.51) (0.55) (0.79) EN1092-1, DIN2501 / 316/316L SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 190 B16.5 / 316/316L (9.57) (11.62) (7.5) (0.75) (0.94) flange 4" 150lb RF, ASME 243.2 295.2 3° 120 (393.7) 228.6 (9) 190.5 (7.5) 19.1 23.9 B16.5 / 316/316L (9.57) (11.62) (0.75) (0.94) SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 191 120 (393.7) 175 (6.89) JIS / 316/316L  (9.57) (11.62) (8.27) (0.75) (0.79) flange DN150 10K RF, 243.2 295.2 3° 120 (393.7) 240 (9.45) JIS / 316/316L (9.57) (11.62) (11.02) (0.91) (0.87) SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 192: Lr550 Polymeric Horn Antenna

    Dimension drawings 13.3 LR550 polymeric horn antenna 13.3 LR550 polymeric horn antenna ① 3" Universal bolted flange - - - - Sensor reference point Dimensions in mm (Inch) SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 193 (2.95) ① ④ Air purge inlet Device flange ② ⑤ Non return valve Process seal ③ Device flange     - - - - Sensor reference point Dimensions in mm (Inch) SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 194 LR550 with mounting bracket LR550 mounting bracket side and top view (4.9") (.47) (4.3") (.12) (4.9) R 4.5 (.17) (3.34) (.47) ø 9 (.35) 122.7 (4.83") (6.25") (16.2) (12.5) (10.7) Dimensions in mm(Inch)   SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 195 22.4 (0.88) DN100 10K FF, JIS 210 (8.27) 175 (6.89) 19 (0.75) DN150 10K FF, JIS 280 (11.02) 240 (9.49) 23 (0.9) Recommended max. range 120 m (393.7 ft) Beam angle 3° SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 196: Lr580 Flanged Lens Antenna

    Dimension drawings 13.4 LR580 flanged lens antenna 13.4 LR580 flanged lens antenna LR580 flat face flanges DImensions in mm (inch)   - - - - Sensor reference point   SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 197 23 (0.07) Aiming flange universal 228.6 (9) 175...191 19.1 (0.06) DN100/4" Aiming flange universal 285 (11.22) 241 (9.49) 24 (0.08) DN150/6" Recommended max. range: 120 m (394 ft) Beam angle: 3° SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 198 LR580 lens antenna with purge connection ‑ ‑ ‑ ‑ Sensor reference point DImensions in mm (inch) LR580 lens antenna with aiming flange and purge connection DImensions in mm (inch)   SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 199: Product Documentation And Support

    – "Download as html5, only PC": Open or save the manual in the HTML5 view on your PC You can also find manuals with the Mobile app at Industry Online Support (https:// support.industry.siemens.com/cs/ww/en/sc/2067). Download the app to your mobile device and scan the device QR code. Product documentation by serial number Using the PIA Life Cycle Portal, you can access the serial number-specific product information including technical specifications, spare parts, calibration data, or factory certificates.
  • Page 200: Technical Support

    (http:// www.siemens.com/automation/csi/service).   Service & support on the Internet In addition to our technical support, Siemens offers comprehensive online services at service & support (http://www.siemens.com/automation/serviceandsupport). Contact If you have further questions about the device, contact your local Siemens representative, by doing the following: 1.
  • Page 201: Technical Reference

    Technical reference Principles of operation SITRANS LR500 series is a 2-wire, 80 GHz FMCW (Frequency Modulated Continuous Wave) radar level transmitter for continuous monitoring of liquids, solids and slurries in vessels to a range of 100 m (329 ft) . Radar level measurement uses the time of flight principle to determine distance to a material surface.
  • Page 202: Echo Selection

    B.2.2.3 Echo selection algorithms The echo is selected based on echo selection algorithms that ultimately use confidence to select the echo. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 203: Loop Power

    Loop power B.3.1 Loop power Note Loop voltage Loop voltage is the voltage at the terminals of the power supply (not the voltage at the terminals of the device). SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 204 Loop voltage - V Note HART communication For example as per graph, when using HART communication with 240 Ohms of Loop resistance ), the minimum Loop voltage (V ) is 16 V DC. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 205: Communication

    The radar device can be configured over the HART network using either the HART Communicator 375 by Fisher-Rosemount, or a software package. The recommended software package is the SIMATIC Process Device Manager (PDM) by Siemens. HART communication protocol This device supports HART communication protocol. Signals are processed using Process Intelligence which has been field-proven in over 1,000,000 applications worldwide (ultrasonic and radar).
  • Page 206 Communication C.4 HART Electronic Device Description (EDD) SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 207: Remote Operation

    Operating Manual 'Help for SIMATIC PDM'. The manual is delivered with SIMATIC PDM software. Once the SIMATIC PDM is installed on your computer you find the manual under: Start > All programs > Siemens Automation > SIMATIC > Documentation. Link at our website: SIMATIC PDM instructions and manuals (https:// support.industry.siemens.com/cs/ww/en/ps/16983/man).
  • Page 208: Check Simatic Pdm Version

    D.2 Bluetooth D.1.3 Check SIMATIC PDM version Procedure 1. Go to SIMATIC PDM Download (http://www.siemens.com/simaticpdm/downloads). 2. Check the support page to make sure you have: – The latest version of SIMATIC PDM – The most recent Service Pack (SP) – The most recent hot fix (HF) D.1.4...
  • Page 209: Connecting Field Device And Sitrans Aw050 Bluetooth Adapter

    The cables are best run separately inside the enclosure as shown in the figure following. Take care when installing the display to be sure that the cables are not pinched or torn. Excessive force should not be necessary for installation of the display mounting. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 210: Installing Or Replacing Bluetooth Adapter

    3. Slide the adapter into the gland against the O-ring. 4. Tighten the gland to the adapter, without twisting the cables. 5. Plug the cables into the display and electronics. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 211: Connecting Field Device With Sitrans Mobile Iq App

    – Do not use the same password for different Bluetooth field devices. – If the mobile end device has device protection enabled, SITRANS mobile IQ automatically saves the password. Individually stored device passwords can be deleted in the app. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 212: Default Password

    Weight   • With cable gland 24 g 13 g • Without cable gland Degree of protection • Type 4X, Type 6 in accordance with UL 50E • IP66, IP68 in accordance with IEC 60529 EN 61326 SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 213: Technical Data: Sitrans Mobile Iq

    CMIIT ID: 2020DJ15120 D.2.7 Technical data: SITRANS mobile IQ Software requirements   Bluetooth BLE 4.2 or higher D.2.8 Dimensions SITRANS AW050 Bluetooth adapter Dimension drawing SITRANS AW050 Bluetooth adapter, dimensions in mm (inch) SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 214: Information For Radio Approvals

    This module can change the output power depending on the circumstances by the application software which is developed by module installer. Any end user cannot change the output power. SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 215: Hart

    SITRANS LR500 can be configured over the HART network using the HART Communicator TREX by Emerson, or a software package. The recommended software package is the SIMATIC Process Device Manager (PDM) by Siemens. HART version SITRANS LR500 conforms to HART revision 7.6.
  • Page 216: Communication Connections

    The SITRANS LR500 can be connected to a computer system via a HART modem (connected to the mA OUT/HART terminal block). Typical PLC/mA configuration with passive HART connection ① ④ Device R= 250 Ω ② ⑤ Power supply HART modem ③ ⑥ HART communicator SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 217: Configuring Communication Ports

    (A custom fixed current value can be entered in Loop current value in multidrop mode (2.4.2) (Page 120).) D.3.5 Communication troubleshooting For more information, refer to Communication troubleshooting (Page 155). SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 218 Remote operation D.3 HART SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 219: Index

    Customer Support, (Refer to Technical support) (EDD), 216 HART Communications details, 205 Hazardous area Laws and directives, 19 damping, 96 Qualified personnel, 23 data logging hazardous area installations device nameplate, 56, time-based echo profile storage, 102 SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 220 Support, 200 Support request, 200 operating principles, 201 Symbol, 158 configuration, 156 device status, 156, 158 diagnostics, 156, 158 Password maintenance, 156, 158 Reset, 212 operating mode, 156 process value, 156 see SIMATIC PDM, 93 Symbols, (Refer to warning symbols) Process values Damping, 74 SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 221 Test certificates, 19 trend view trend line, 106 troubleshooting communication, 165 TVT Shaper manual adjustment via PDM, 101 via PDM, 102 Volume calculation example, 89 vessel shape, 78, 123 Warning symbols, 19 Warranty, 16 wear view via PDM, 105 SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...
  • Page 222 Index SITRANS LR500 series with mA/HART Operating Instructions, 03/2024, A5E51099898-AA...

This manual is also suitable for:

Sitrans lr510Sitrans lr550Sitrans lr530Sitrans lr580

Table of Contents