Enclosure Open - Siemens SITRANS LR500 Series Operating Instructions Manual

Radar level transmitters
Hide thumbs Also See for SITRANS LR500 Series:
Table of Contents

Advertisement

10.3.1

Enclosure open

Enclosure open
Risk of explosion in hazardous areas.
To open the device in a hazardous area, isolate the device from power.
Exception: Devices exclusively having Intrinsic safety (Ex i) may be opened in an energized
state in hazardous areas.
Improper connection after maintenance
Risk of explosion in areas subject to explosion hazard.
• Connect the device correctly after maintenance.
• Close the device after maintenance work.
Refer to Connecting (Page 49).
Hot, toxic, or corrosive process media
Risk of injury during maintenance work.
When working on the process connection, hot, toxic, or corrosive process media could be
released.
• As long as the device is under pressure, do not loosen process connections and do not
• Before opening or removing the device ensure that process media cannot be released.
Hot surfaces
Risk of burns during maintenance work on parts having surface temperatures exceeding 70 °C
(158 °F).
• Take corresponding protective measures, for example, by wearing protective gloves.
• After carrying out maintenance, remount touch protection measures.
SITRANS LR500 series with mA/HART
Operating Instructions, 03/2024, A5E51099898-AA
WARNING
WARNING
WARNING
remove any parts that are pressurized.
CAUTION
Service and maintenance
10.3 Maintenance and repair work
151

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sitrans lr510Sitrans lr550Sitrans lr530Sitrans lr580

Table of Contents