Digital compressive strength tester
Hide thumbs Also See for Chandler Engineering 4207D:

Advertisement

Quick Links

OPERATING MANUAL
MODEL 4207D
DIGITAL COMPRESSIVE
STRENGTH TESTER
Revision M – November 2015
P/N: 89-0164
S/N: ____________
:
"
"
: 614007, .
,
. 25
72,
40
: +7 (342) 262-85-56
: +7 (342) 262-85-60
email: office@techimport.ru
www.techimport.ru

Advertisement

Table of Contents
loading

Summary of Contents for Ametek Chandler Engineering 4207D

  • Page 1 OPERATING MANUAL MODEL 4207D DIGITAL COMPRESSIVE STRENGTH TESTER Revision M – November 2015 P/N: 89-0164 S/N: ____________ “ ” : 614007, . . 25 : +7 (342) 262-85-56 : +7 (342) 262-85-60 email: office@techimport.ru www.techimport.ru...
  • Page 2 This publication contains the following trademarks and/or registered trademarks: AMETEK, CHANDLER ENGINEERING. These trademarks or registered trademarks and stylized logos are all owned by AMETEK, Inc. All other company, product and service names and logos are trademarks or service marks of their respective owners.
  • Page 3: Table Of Contents

    TABLE OF CONTENTS Table of Contents General Information ..............P-1 Introduction..........................P-1 Purpose and Use ....................P-1 Description ......................P-1 Features and Benefits ......................P-1 Specifications ......................... P-1 Safety Requirements ......................P-2 Safety Features ........................P-2 Section 1 – Installation ............. 1-1 Unpacking the System ......................
  • Page 4 TABLE OF CONTENTS Section 4 - Troubleshooting Guide .......... 4-1 Section 5 - Replacement Parts ..........5-1 Section 6 – Drawings and Schematics ........6-1...
  • Page 5: General Information

    PREFACE General Information Introduction The Model 4207D Digital Compressive Strength Tester is a hydraulic press system that may be used to apply known compressive loads to a sample at known rates. The maximum load is 50,000 Lbf. The Model 4207D Tester meets all the requirements for cement compressive testing as specified in API Specification 10.
  • Page 6: Safety Requirements

    PREFACE Serial Interface: Modbus-RTU Protocol Hydraulic Fluid: SAE 10W30 Synthetic Oil Shipping Dimensions: Load Frame 48” (122 cm) high x 24” (61 cm) wide x 28” (71 cm) deep Control Cabinet 54” (138 cm) high x 28” (71 cm) wide x 30” (76 cm) deep Net Weight: Load Frame 360 lbs (164 kg)
  • Page 7 PREFACE • Automatic system shut-down: The controller automatically terminates the active program when the sample fails.
  • Page 8 PREFACE This page is intentionally left blank...
  • Page 9: Section 1 - Installation

    SECTION 1 – INSTALLATION Section 1 – Installation Unpacking the System Carefully unpack the Model 4207D and all of its accessories. Visually inspect for any damage that may have occurred during shipping. After the instrument is removed from the shipping crate, the equipment and spare parts should be checked against the packing list to insure that all parts have been received and none are damaged.
  • Page 10: Connecting The Press To The Power Unit

    SECTION 1 – INSTALLATION The system power switch on the front panel also serves as a circuit breaker. If the breaker trips, correct the electrical problem then reset the breaker by cycling the switch. Connecting the Press to the Power Unit There are three connections between the load frame and the control cabinet, 2 hydraulic lines and a 25-pin cable connection.
  • Page 11: Section 2 - Operating Instructions

    SECTION 2 – OPERATING INSTRUCTIONS Section 2 - Operating Instructions Before operating the 4207D it is necessary to be familiar with the press controller and load indicator features and controls. Programming the Controller The Model 7052 controller used with the 4207D system features user defined segment programming (8 segments maximum).
  • Page 12: Sample Controller Program

    SECTION 2 – OPERATING INSTRUCTIONS Configuring the controller to perform a sample load program involves defining a series of ramp and dwell segments. Once the program exists, the program is executed by pressing the Run button. To suspend the program the Run/Hold button may be pressed briefly (Hold light illuminates) and restarted by pressing the Run/Hold button again.
  • Page 13: Configuring The Display

    SECTION 2 – OPERATING INSTRUCTIONS Segment Segment Value, Time, min Comment Number Type Lbf/10 Ramp Time This is used to load the sample. Dwell Ramp Time 4800 3.00 or 12.00 API Spec 10A Ramp rates of 16,000 Lbf/min or 4,000 Lbf/min. Rset Ends the program once the setpoint is reached.
  • Page 14: Operating Procedure

    SECTION 2 – OPERATING INSTRUCTIONS calibration values must be configured in the display and controller (CONFIG – IP menus). The display is configured to display the current value. To display the peak value, press the UP or DOWN arrow, the peak value will display for 2 seconds then revert to the current value.
  • Page 15 SECTION 2 – OPERATING INSTRUCTIONS 7. If using a data acquisition system, start the program. 8. Confirm the controller is idle and in manual (MAN) mode by performing the steps below. a. Press and hold the RUN/HOLD button for three seconds or longer. This will reset and idle the controller.
  • Page 16: Interpreting The Results

    SECTION 2 – OPERATING INSTRUCTIONS a. Press the RUN/HOLD button. The RUN indicator will illuminate. b. The program will increase the applied force to 150 Lbf (15 x 10) clamping force in 30 seconds. c. The program will hold the 150 Lbf clamp force for another 30 seconds. d.
  • Page 17 SECTION 2 – OPERATING INSTRUCTIONS Maximum Load Compressiv Strength Sample CrossSecti onal Area If a standard ASTM cement cube is used, the cross-sectional area is 4.0 in (2 inch x 2 inch cube). To obtain the Compressive Strength in psi, divide the Maximum Load value by 4.0.
  • Page 18 SECTION 2 – OPERATING INSTRUCTIONS This page is intentionally left blank...
  • Page 19: Section 3 - Maintenance

    SECTION 3 – MAINTENANCE Section 3 – Maintenance Maintenance Schedule MAINTENANCE SCHEDULE Model 4207D Digital Compressive Strength Tester COMPONENT EACH TEST MONTHLY 3 MONTHS 6 MONTHS ANNUAL Check Level Every two years, change the oil System Pressure Check the system pressure, adjust to 2600 psi Press Assembly...
  • Page 20: Resetting The Maximum Temperature Limit

    SECTION 3 – MAINTENANCE Note: The system pressure will decrease as the oil temperature increases. For this reason, adjust the system pressure when the oil temperature is at or near room temperature. 1. Verify that the system temperature is below 80°F (27°C). Gauge 2.
  • Page 21: Maintaining The Press Assembly

    SECTION 3 – MAINTENANCE Maintaining the Press Assembly Use the following procedure to inspect the condition and safety of the press assembly: 1. Verify that the door limit switch is operational. 2. Inspect the cylinder and repair any leaks. 3. Inspect the hydraulic hoses and replace if signs of deterioration exist. 4.
  • Page 22: Calibration

    SECTION 3 – MAINTENANCE The limit switch may require periodic adjustment. Use the following procedure to adjust the limit switch: Warning: Make sure the instrument power is off before performing this procedure. 1. Remove the cover plate from the limit switch assembly. 2.
  • Page 23: Controller Calibration

    SECTION 3 – MAINTENANCE 5. Update the calibration factor used in the display and controller. Follow the manufacturers instructions for updating the factor. (CONFIG – IP menus). Controller Calibration The controller uses the same load cell signal that is used by the display. For this reason, they use the same calibration factor.
  • Page 24 SECTION 3 – MAINTENANCE This page is intentionally left blank...
  • Page 25 SECTION 4 - TROUBLESHOOTING GUIDE Section 4 - Troubleshooting Guide Symptom Reason Action Press does not operate. Load frame door is open. Close the load frame door. Over-temperature circuit has Reset the over-temperature disabled the system. circuit by pressing the red button on the side panel.
  • Page 26 SECTION 4 – TROUBLESHOOTING GUIDE Symptom Reason Action With zero load, adjust the OFS.1 offset value in the controller IP menu until the value agrees with the load display. Control program does not end Controller “rate of change” set Set the set point value for alarm when a cube fails.
  • Page 27 SECTION 5 - REPLACEMENT PARTS Section 5 - Replacement Parts Part Number Description 07-0176 Thermocouple Assembly 7052 Controller 89-0022 Base, Painted 89-0150 Panel, Front 89-0155 Mandrel, Top, Load Cell 89-0156 Mandrel - Bottom - Load Cell 89-0181 Power Cable Assembly 89-0190 Load Cell Assembly 89-0191...
  • Page 28 SECTION 5 – REPLACEMENT PARTS This page is intentionally left blank...
  • Page 29 SECTION 6 - DRAWINGS AND SCHEMATICS Section 6 – Drawings and Schematics Document Number Description Controller Configuration 89-0151 Wiring Diagram 89-0170 4207D Pump Connections 89-0175 Procedure, 5270 Configuration...
  • Page 30 SECTION 6 - DRAWINGS AND SCHEMATICS This page is intentionally left blank...
  • Page 31 CONTROLLER CONFIGURATION Controller Configuration The controller chassis will slide out of the back shell by loosening the clips at the right and left front of the controller front panel. The controller is installed in the panel of the instrument using two plastic clips located behind the panel.
  • Page 32 CONTROLLER CONFIGURATION Model 4207D Controller Configuration Parameters Parameter Description Status Value Program Run List Hide PROG Program Edit List Altr Holdback type Hide Hb.U Holdback units Hide rmP.U Ramp units Hide dwL.U Dwell units Hide CYC.n Number of program Hide cycles SEG.n Segment number...
  • Page 33 CONTROLLER CONFIGURATION Model 4207D Controller Configuration Parameters Parameter Description Status Value OFS.1 Lil.2 not used Hi.i not used not used PU.iP iP.1 iP.1 Factory/User Cal FACT CAL.S calibration point none not used OFS.1 IP1 cal offset Varies with zero load output of load cell OFS.2 not used...
  • Page 34 CONTROLLER CONFIGURATION Model 4207D Controller Configuration Parameters Parameter Description Status Value dTyp Front panel Auto/Man EnAb Front panel Run/Hold EnAb PwrF Power feedback Fwd.t Feed forward type None Pdtr Manual/Auto transfer Sbr.t Sensor break output Sb.OP Forced manual output Step BCD input None GSch...
  • Page 35 CONTROLLER CONFIGURATION Model 4207D Controller Configuration Parameters Parameter Description Status Value Srvo Starting setpoint of a to.SP program Digital Input 1 Configuration Identity LoG.i Func Function of input Stby Digital Input 2 Configuration Identity LoG.i Func Function of input Comms 1 Configuration Identity of module Func Function...
  • Page 36 CONTROLLER CONFIGURATION Model 4207D Controller Configuration Parameters Parameter Description Status Value Alarm 1 Alarm 1 active Alarm 2 Alarm 2 active Alarm 3 Alarm 3 active Alarm 4 Rate of change alarm Controller in manual mode Sensor break SPAn PV out of range Loop break Ld.F Load failure alarm...
  • Page 37 CONTROLLER CONFIGURATION Display Configuration The following table contains the default configuration for the load cell display. FULL Comment CONFIG Comment MENUS MENUS HOME INST dSP.F Displays the present value unit None No Engineering units on the front display dSP.b PV.Hi Displays the peak value on dEc.P nnnn.
  • Page 38 CONTROLLER CONFIGURATION FULL Comment CONFIG Comment MENUS MENUS nw.AL Communication module installed Func Modbus protocol bAud 9600 Baud rate PrtY Parity dEL.Y None Response delay Full Resolution Not used Func None VAL.L VAL.H SG.SU Strain gage power supply installed Func IP.1 Bridge supply for input 1 brG.U...
  • Page 39 CHANDLER ENGINEERING...
  • Page 41 NOTE: The latest version of the 5270 DACS software must be used for all features to be supported. At this writing, the most recent version of 5270 is 1.10.097. If an upgrade is needed, please contact AMETEK Chandler Engineering at 918.250.7200 Using Tools –...
  • Page 42 CHANDLER ENGINEERING Part Number: 89-0175 Page 2 of 7 TITLE: Procedure, 5270 Configuration for 4207D Instrument Model: Model 4207D Compressive Strength Tester Using Tools – Configure – Instruments, select the Model 4207D Compressive Strength Tester from the list of available instruments. Assign a name and file prefix.
  • Page 43 CHANDLER ENGINEERING Part Number: 89-0175 Page 3 of 7 TITLE: Procedure, 5270 Configuration for 4207D Instrument Model: Model 4207D Compressive Strength Tester Define the signal I/O addresses as shown. Note that the address for the controller (Load – Programmed) is 1:2f to acquire the programmed load, the address for the peak value (Load –...
  • Page 44 CHANDLER ENGINEERING Part Number: 89-0175 Page 4 of 7 TITLE: Procedure, 5270 Configuration for 4207D Instrument Model: Model 4207D Compressive Strength Tester Once this is complete, the Load – Maximum signal must be configured to provide automatic reset of the peak value when a test is started on 5270.
  • Page 45 CHANDLER ENGINEERING Part Number: 89-0175 Page 5 of 7 TITLE: Procedure, 5270 Configuration for 4207D Instrument Model: Model 4207D Compressive Strength Tester Select the Load – Maximum signal as illustrated. Select Modify.
  • Page 46 CHANDLER ENGINEERING Part Number: 89-0175 Page 6 of 7 TITLE: Procedure, 5270 Configuration for 4207D Instrument Model: Model 4207D Compressive Strength Tester Select the Advanced option. Change the Setpoint 1 and Setpoint 2 values to 2:133f as shown. This action directs 5270 to download a 0.00 value to the display when a test is started.
  • Page 47 CHANDLER ENGINEERING Part Number: 89-0175 Page 7 of 7 TITLE: Procedure, 5270 Configuration for 4207D Instrument Model: Model 4207D Compressive Strength Tester When a test is started using 5270, the following display appears (disregard the RED signal indications, on an actual instrument they must be GREEN) Select the desired API loading rate from the pre-defined test profiles.