ESSENTIAL HEALTH and SAFETY WDG-V HP Series and WDG-V/S Series Combustion Analyzers Class I and Class II, Division 2, ATEX, IECEx Zone 2 PN 10686-17-5_Rev A...
Page 2
AMETEK Process Instruments assumes no responsibility or liability for any errors or inaccuracies that may appear in this document. AMETEK Process Instruments is not responsible for any infringement of patents or other rights of third parties that may result from the use of this document. The content of this document is furnished for informational purposes only, is subject to change without notice, and does not represent a commitment or guaranty by AMETEK Process Instruments.
WDG-V/S Series Analyzer Specifications ..................15 APPROVALS AND CERTIFICATIONS WDG-V HP Series and WDG-V/S Series Analyzer Directives and Standards ....... 17 Remote Calibration Unit (RCU) Certifications/Approvals ............17 WDG-V HP Analyzer, Rapid Purge Labels, 115 VAC and 230 VAC ..........18 WDG-V HP Analyzer, Standard Purge Labels, 115 VAC and 230 VAC ........
Page 4
Installing the Standard 1/8-inch NPT Sample Probe (WDG-V/S Series) ....35 Installing the Ceramic Probe or Long Probe (WDG-V/S Series) ......36 Installing the Exhaust Tube ....................37 Installing the WDG-V HP Series and WDG-V/S Series Analyzer ........38 Installing the WDG-V HP Series Analyzer ................38 Analyzer Flush-to-Process Installation for WDG-V HP Series (Preferred Method) ......................38...
Page 5
AMEVision Connections ........................ 68 Remote Calibration Unit Connections ..................68 Oxygen-Only Remote Calibration Unit Connections ............68 Combustibles Remote Calibration Unit Connections .............68 Methane Remote Calibration Unit Connections ..............68 Digital Input to Initiate Remote Calibration Unit .............69 Notes About Ducting ..........................70 Ducting Between Pressurized Enclosure and Inlet ..............
Page 6
This page intentionally left blank. ® iv | Thermox WDG-V HP Series and WDG-V/S Series Analyzers PN 10686-17-5, Rev A...
About This Document This document describes the Essential Health and Safety requirements (EH&S) for the WDG-V HP Series and WDG-V/S Series (see Note) Combustion Analyzers intended for use in Class I and Class II, Division 2, ATEX, IECEx Zone 2 explosive atmospheres and hazardous locations.
An operating procedure which, if not strictly observed, may result in damage to the equipment. Important information that should not be overlooked. Burn hazard. Hot surface. Do not touch, allow to cool before servicing. 2 | Thermox ® WDG-V HP Series and WDG-V/S Series Analyzers PN 10686-17-5, Rev A...
Warnings and Cautions Before installing, operating, and performing any maintenance, service, and/or troubleshooting on the analyzer, read and follow all safety information in this document (and in the analyzer User Manu- al). Misuse of this product could result in personal injury and/or dam- age to the equipment or its installation site.
Page 10
For electric-shock protection, the analyzer must be operated from a grounded power source that has a securely connected protective- ground contact. 4 | Thermox ® WDG-V HP Series and WDG-V/S Series Analyzers PN 10686-17-5, Rev A...
Warning Labels These symbols may appear on the instrument to alert you of existing condi- tions. Protective Conductor Terminal (BORNIER DE L’ECRAN DE PROTECTION) Schutzerde Caution – Risk of electric shock (ATTENTION – RISQUE DE DÉCHARGE ÉLECTRIQUE) Achtung – Hochspannung Lebensgefahr Caution –...
Environmental Information (WEEE) This AMETEK product contains materials that can be reclaimed and recycled. In some cases the product may contain materials known to be hazardous to the environment or human health. In order to prevent the release of harmful sub-...
Special Conditions for Safe Use The WDG-V HP Series and WDG-V/S Series Analyzers are designed to be used in Category 3 hazardous locations where explosive gases may be present. Protec- tion against explosion is provided via enclosure purging and pressurization.
However, some electronic devices may be susceptible to interference, therefore, ensure that this equipment is used in a suitable environment. The various configurations of the WDG-V HP Series and WDG-V/S Series Analyzers should not produce, or fall victim to, electromagnetic disturbances as specified in the European Union’s EMC Directive.
Supplier’s Declaration of Conformity 47 CFR § 2.1077 Compliance Information Unique Identifier: WDG-V HP Series and WDG-V/S Series Combustion Analyzers Responsible Party – U.S. Contact Information: AMETEK Process Instruments 150 Freeport Road Pittsburgh, Pennsylvania, USA 15238 Tel: 1-412-828-9040 ametekpi.com FCC Compliance Statement (WDG-V HP Series and WDG-V/S Series Com- bustion Analyzers subject to Part 15): This device complies with Part 15 of the FCC Rules.
All input and output wiring must be in accordance with wiring meth- ods authorized for the area classification by the authority having jurisdiction and IEC 60079-14. 10 | Thermox ® WDG-V HP Series and WDG-V/S Series Analyzers PN 10686-17-5, Rev A...
About the Analyzer The WDG-V HP Series Combustion Analyzer is a flue gas analyzer, designed for high particulate flue gas applications. The analyzer uses convection as a mo- tive force and mounts directly to the combustion process to provide continu-...
Page 18
This page intentionally left blank. 12 | Thermox ® WDG-V HP Series and WDG-V/S Series Analyzers PN 10686-17-5, Rev A...
Specifications WDG-V HP Series Analyzer Specifications For the purpose of this document, specifications listed here apply primarily to the utilities required for the WDG-V HP Series Combustion Analyzer User Manual. Output Range Oxygen: From 0–1 % to 0–100 % Combustibles: From 0–2,000 PPM (standard)
Page 20
Hinged NEMA 4X (IP65). NOTES All static performance characteristics are with operating variables constant. System accuracy referenced to 0.1–10 % calibrated range. Response is to calibration gas. 14 | Thermox ® WDG-V HP Series and WDG-V/S Series Analyzers PN 10686-17-5, Rev A...
WDG-V/S Series Analyzer Specifications For the purpose of this document, specifications listed here apply primarily to the utilities required for the WDG-V/S Series Combustion Analyzer User Manual. Output Range Oxygen: From 0–1 % to 0–100 % Combustibles: From 0–2,000 PPM (standard) (optional 0–10,000 PPM, 0–1 %, 0–5 %) Methane: From 0–1 % to 0–5 %...
Page 22
Ambient Temperature: -18 °C to 60 °C (0 °F to 140 °F) Ambient Humidity: 0 % to 90 %, non-condensing Maximum Altitude: 2000 meters IEC Installation: Category II IEC Pollution Degree 2 16 | Thermox ® WDG-V HP Series and WDG-V/S Series Analyzers PN 10686-17-5, Rev A...
Approvals and Certifications WDG-V HP Series and WDG-V/S Series Analyzer Directives and Standards IECEx IECEx ETL 23.0047X: IEC 60079-0: 2017, Edition 7.0 IEC 60079-2: 2014, 6th Edition ATEX ETL23ATEX0334X: Directive 2014/34/EU using the following standards: EN IEC 60079-0: 2018 EN 60079-2: 2014...
WDG-V HP Series Analyzer, North American Rating Labels, 115 VAC and 230 VAC 22 | Thermox ® WDG-V HP Series and WDG-V/S Series Analyzers PN 10686-17-5, Rev A...
Analyzer Purge Markings The analyzer component layout drawings shown in this section il- lustrate a WDG-V HP Series Analyzer. The labels and their locations are located similarly on a WDG-V/S Series Analyzer. WDG-V HP Series and WDG-V/S Series Rapid Purge...
Page 31
Figure 1. Rapid Purge system, WDG-V HP Series shown. PN 10686-17-5, Rev A EH&S...
WDG-V HP Series and WDG-V/S Series Standard Purge Warning Label and Component Layout Drawings 26 | Thermox ® WDG-V HP Series and WDG-V/S Series Analyzers PN 10686-17-5, Rev A...
Page 33
Figure 2. Standard Purge system, WDG-V HP Series shown. PN 10686-17-5, Rev A EH&S...
Page 34
Figure 3. Optional external pressure switch, standard WDG-V HP Series Purged Analyzer shown. 28 | Thermox ® WDG-V HP Series and WDG-V/S Series Analyzers PN 10686-17-5, Rev A...
Analyzers. Refer service requirements to qualified personnel. Read this entire section before beginning installation of the analyzer. Failure to do so, and/or use of the WDG-V HP Series or WDG-V/S Series Analyzer in a manner not specified in this manual, may impair the protection against fire, electrical shock, and personal injury originally provided by this equipment.
Open the Remote Calibration Unit (RCU) Enclosure and inspect all interior components for obvious damage and broken/loose parts or fittings. If equipment is damaged, notify the carrier and contact AMETEK Service (https://www.ametekpi.com/customersupport/requestsupport) immediately if parts are missing or damage is found, and to verify if damaged parts will re- quire replacement prior to safely installing and operating the analyzer/equip- ment.
Analyzer Site Preparation The WDG-V HP Series and WDG-V/S Series Analyzers are approved for installa- tion in hazardous areas classified as Class I, Class II, Division 2, Zone 2. Observe the following guidelines when selecting an analyzer installation location: •...
Always install the analyzer in a location that is as free as possible from vibrations. • The WDG-V/S Series and WDG-V HP Series Analyzers are rated for outdoor operation (IP65 / NEMA 4X and ambient -20 °C to +60 °C (-4 °F to 140 °F). 32 | Thermox ®...
Installing the Sample Inlet Probe This section includes separate procedures on how to install the Sample Inlet Probe for WDG-V HP Series and WDG-V/S Series Analyzers. Installing the Sample Inlet Probe and Probe Filter (WDG-V HP Series) The sample probe and filter must be attached to the back of the sensor before mounting the sensor to the process.
In that case you should mount the analyzer first and then install the probe. 34 | Thermox ® WDG-V HP Series and WDG-V/S Series Analyzers PN 10686-17-5, Rev A...
Installing the Standard 1/8-inch NPT Sample Probe (WDG-V/S Series) Connect a standard 1/8-inch NPT Sample Probe to the analyzer as follows: • If your probe is 4 feet or less, simply thread the probe into the back of the sample probe port on the rear of the analyzer (Figure 7), then install the analyzer.
4. Thread the ceramic or long probe into the sensor inlet port on the sensor backplate. Do not let the probe hit against the AMETEK-provided pipe nip- ple as you insert the probe into the process stream. If inserting a ceramic probe, you must slowly insert it into a hot process stream to prevent the probe from cracking due to thermal shock.
If installing an exhaust tube, you must connect it to the analyzer be- fore mounting the analyzer to the process stream. An optional stainless steel exhaust tube (AMETEK Part No. 70619KE) is available if an extension to the aspirator exhaust is needed.
Enclosure door until you first verify that the area has been classified as nonhazardous. There are three (3) different methods to mount a WDG-V HP Series Analyzer to the process gas. Each of these methods is explained in the following pages.
Page 45
3. Weld the AMETEK-supplied mounting plate to the process wall. 4. Mount the analyzer to the process as described in the “Mounting the Sen- sor to the Process” section later in this chapter. This method positions the analyzer flush with the process and helps to ensure that the sample gas does not cool below its dewpoint.
1. Weld a short 3-1/2" diameter pipe nipple to the duct and extend through the wall. The shorter the pipe nipple, the better. 2. Insert the AMETEK-supplied pipe nipple (on the end of the mounting plate) into your pipe nipple. If the pipe nipple extends past the inside of the refractory wall on a process which has a gas temperature over 650 °C,...
WDG-V HP Series Mount the analyzer to a 3" x 300# customer-supplied flange (Figure 10). In this case, you will not need the AMETEK-supplied mounting plate with attached protective pipe nipple. Mount the analyzer to the process as described in the “Mounting the Sensor to the Process”...
Analyzer Flush-to-Process Installation for WDG-V/S Series (Preferred Method) Weld the 7" x 7-5/8" (17.78 cm x 19.37 cm) mounting plate (AMETEK Part No. 70626SE) supplied with the analyzer to the process wall over a 3-3/4" (9.52 cm) diameter hole (Figure 11). Then, bolt the analyzer to the mounting plate. This method positions the analyzer as close to the process as possible and mini- mizes any chance that the sample gas will cool below its dewpoint.
Mount the analyzer to a 3" x 300# customer-supplied flange (Figure 12). In this case, you will not need the AMETEK-supplied mounting plate with attached protective pipe nipple. If you have a different type of flange, AMETEK supplies a number of flange adapters.
3. Wrap the nipple with at least 1" (2.54 cm) weatherproof insulation and heat trace it. 4. Bolt the analyzer to the plate. Figure 13. WDG-V/S Series Analyzer customer-supplied pipe nipple installation. 44 | Thermox ® WDG-V HP Series and WDG-V/S Series Analyzers PN 10686-17-5, Rev A...
1. Weld a customer-supplied 3" x 150# flange and nipple to the top of the duct (Figure 14). 2. Bolt the analyzer assembly in place using AMETEK-supplied hardware and gasket. 3. Wrap the nipple with at least 1" weatherproof insulation around exposed portion.
Page 52
Figure 15. WDG-V/S mounting dimensions, hinged door analyzers. A 7/16-inch wrench is required to open both doors to the sensor. 46 | Thermox ® WDG-V HP Series and WDG-V/S Series Analyzers PN 10686-17-5, Rev A...
Purge Connection Options For North American Class I Division 2 and ATEX/IECEx Zone 2 applica- tions, AMETEK provided two purge options. Standard Purge Configuration • Connect instrument air to the analyzer Z-purge inlet (Figure 16). Maintain the pressure and flow as directed on the Z-purge warning label.
Page 54
No filters shall be fitted between the pressurized enclosure and the protec- tive gas system alarm. • No isolating valve shall be fitted between the indicator and the pressurized enclosure. Figure 16. Standard purge component layout. 48 | Thermox ® WDG-V HP Series and WDG-V/S Series Analyzers PN 10686-17-5, Rev A...
65 minutes at a pressure of 0.4” of water. Always use tubing that is free of oil and dirt. Stainless steel tubing is recom- mended. SUPPLIED BY SUPPLIED BY CUSTOMER AMETEK MARKED SERVICE VALVE ADJACENT TO ENCLOSURE GAUGE ALARM PRESSURIZED...
Refer to Figure 16 for details. 5. Carefully remove T-bar Valve Key from the Enclosure Pressure Control Valve stem. Ensure the Enclosure Pressure Gauge “Safe” pressure setting is stable. 50 | Thermox ® WDG-V HP Series and WDG-V/S Series Analyzers PN 10686-17-5, Rev A...
Page 57
Use the T-bar Valve Key to open the Rapid Exchange control Valve fully by turning 90° counter-clockwise, and quickly ensure the Enclosure Protec- tion Vent opens. The Enclosure Pressure Gauge should move quickly off scale to the right. This is normal for all Rapid Exchange purging systems. Observe the system during the exchange time duration (five minutes mini- mum), then close the Rapid Exchange Control Valve fully and ensure the T-bar Valve Key is firmly locked in the Rapid Exchange Control Valve stem.
– note the orientation of the valve. 5. Connect the aspirator air outlet on the RCU to the aspirator air inlet con- nection on the analyzer. 52 | Thermox ® WDG-V HP Series and WDG-V/S Series Analyzers PN 10686-17-5, Rev A...
Page 59
Figure 18. RCU mounting dimensions, version shown (1 span). PN 10686-17-5, Rev A EH&S...
Page 60
Figure 19. Flow diagram, standard RCU, version shown (1 span). 54 | Thermox ® WDG-V HP Series and WDG-V/S Series Analyzers PN 10686-17-5, Rev A...
Combustibles RCU Installation Figure 20 shows mounting dimensions for the combustibles RCU. The ambient temperature range for the RCU is -20° to 70°C. Mount the RCU as close to the analyzer as possible. Shorter calibration plumbing improves response times, reduces calibration gas expense, and reduces the chance of contaminants in the calibration gas plumbing.
Page 62
Figure 20. RCU mounting dimensions, Combustibles. 56 | Thermox ® WDG-V HP Series and WDG-V/S Series Analyzers PN 10686-17-5, Rev A...
Installing the AMEVision Control Unit The AMEVision Control Unit can be mounted to either a panel or a wall. Refer to the AMEVision Display User Interface User Manual (AMETEK Part No. 9000- 165-VE). Figure 23. AMEVision mounting dimensions. 58 | Thermox ®...
AC Power and Signal Connections Remove AC mains power from the AMEVision before beginning wiring. Connections to the analyzer are made through the Wiring board. The Wiring board is located on the right side of the analyzer. Any screw terminals on the Wiring board not described in this section are reserved for future use.
Page 66
Figure 24. Conduit Entries. 60 | Thermox ® WDG-V HP Series and WDG-V/S Series Analyzers PN 10686-17-5, Rev A...
• Chassis Stud – Equipment ground (protective conductor) Use the 1/2-inch conduit entry hole in the WDG-V HP Series and WDG-V/S Series analyzer for AC mains supply wiring. Use the chassis stud next to the 1/2-inch conduit entry hole for equipment ground (protective conductor).
Keep shields as short as possible. Ground Stud Method • Connect all cable shields for that conduit entry to the grounding stud clos- est to that conduit entry hole. 62 | Thermox ® WDG-V HP Series and WDG-V/S Series Analyzers PN 10686-17-5, Rev A...
Page 69
Figure 25. Direct shield grounding methods. PN 10686-17-5, Rev A EH&S...
As an example, do not run motors, blowers, or air conditioners using the same mains supply circuit or conduit as the analyzer mains supply circuit or conduit. 64 | Thermox ® WDG-V HP Series and WDG-V/S Series Analyzers PN 10686-17-5, Rev A...
Current Output Connections Standard Current Outputs There are three (3) current outputs on the WDG-V HP Series and WDG-V/S Series Analyzers. The current output connections are labeled as follows on the analyzer board terminals: Analog Output #1 => +I1- Analog Output #2 => +I2- Analog Output #3 =>...
Alarm Contact Connections Standard Alarm Connections The WDG-V HP Series and WDG-V/S Series Analyzers provides five (5) sets of normally open alarm contacts as follows: Terminal Block ID Description – Service Alarm – Data Valid Alarm ALM1 – Process Alarm 1 (Configurable) ALM2 –...
If you have the methane option, you must also make the following connec- tions: - METHANE SPAN – WDG-V Terminal S3 to Pin 18 on RCU 68 | Thermox ® WDG-V HP Series and WDG-V/S Series Analyzers PN 10686-17-5, Rev A...
Figure 27. RCU connections. Digital Input to Initiate Remote Calibration Unit The digital input connections on the Wiring board allow you to initiate a remote calibration from a remote location. For this option to work, you must have a remote calibration unit. The system monitors the digital input, and when the switch closes, it sends the system into an automatic calibration.
Enclosure Maximum Overpressure The user should limit the pressure as specified by the manufacturer. 70 | Thermox ® WDG-V HP Series and WDG-V/S Series Analyzers PN 10686-17-5, Rev A...
System Start-Up and Verification When powering up the analyzer, certain checks should be performed to en- sure the analyzer is operating properly and is ready to be commissioned into operation. Resolve any problems before proceeding to the next start-up stage. Before proceeding, test the area around the analyzer for flammable gases and proceed only when the area is found to be safe.
WDG-V HP Series and the WDG-V/S Series Analyzers. For North American Class I Division 2 and ATEX/IECEx Zone 2 applica- tions, AMETEK provides two purge options (Standard Purge and Rapid Purge) that can be used with the WDG-V HP Series and WDG-V/S Series Analyzers. Standard Purge Start-Up...
Page 79
Figure 28. Standard Purge system, WDG-V HP Series shown. PN 10686-17-5, Rev A EH&S...
9. Close the rapid exchange control and furnace purge valves. 10. Ensure that a safe pressure is maintained. 11. Energize the analyzer. 12. Pressure loss requires immediate attention when the unit is energized. 74 | Thermox ® WDG-V HP Series and WDG-V/S Series Analyzers PN 10686-17-5, Rev A...
Page 81
Figure 29. Rapid Purge system, WDG-V HP Series shown. PN 10686-17-5, Rev A EH&S...
Page 82
WITH INDICATOR ALARM PROJECTIVE ACTUATOR SUPPLY MARKED REGULATOR WGD-V/HP Figure 30. ADJACENT TO ENCLOSURE Standard purge setup. Figure 31. Optional external pressure switch on Purged Analyzer. 76 | Thermox ® WDG-V HP Series and WDG-V/S Series Analyzers PN 10686-17-5, Rev A...
To clean and/or replace parts in the various assemblies that make up the sample system analyzer: The WDG-V HP Series and WDG-V/S Series Analyzers are available in two (2) operating voltage configurations (115 VAC and 230 VAC), con- figured by AMETEK at the factory per customer/end user requirements.
Adjusting the Setpoint To change setpoint: • Remove the plastic cap and turn the slotted Adjustment Screw at the top of the housing clockwise to raise the setpoint pressure and counter-clock- wise to lower the setpoint. After calibration, replace the plastic cap and re-check the setpoint.
Page 88
82 | Thermox ® WDG-V HP Series and WDG-V/S Series Analyzers PN 10686-17-5, Rev A...
Page 90
This page intentionally left blank. 84 | Thermox ® WDG-V HP Series and WDG-V/S Series Analyzers PN 10686-17-5, Rev A...
Page 91
AMETEK Process Instruments delivers worldwide sales and service support through a network of direct and factory-trained global distribution channels. AMETEK Service Assistance Program plans offer coverage up to 24 hours a day, 365 days of the year. As worldwide experts in the manufacture of process analyzers and instrumentation, we have supplied solutions to industry since 1962, providing the widest range of analysis technology available.