Siemens SITRANS TDL Compact Operating Instructions page 19

In situ laser gas analyzers
Hide thumbs Also See for SITRANS TDL:
Table of Contents

Advertisement

3° or 1° wedge window
standard version
1.
Put graphite gasket on the purging tube and insert
into process flange.
2.
Check if the window in flange is clean. Look through
the window on a source of white light or its surround-
ings. If there is dust or dirt, clean with a dry, dustless
cloth and then purge with dry air. Isopropyl alcohol
may also be used.
3.
Mount flange on the pro-
cess flange with another
graphite gasket between
flange and purging tube.
Screw on M16 nuts with
pads and tighten them
with 14 mm hex key and
24 mm wrench. Tighte-
ned nuts with torque (ap-
proximately 60 ... 70 Nm)
for good seal on the gas-
kets.
SITRANS TDL
Compact Operating Instructions, 05\2023, A5E51659183-AA – revised 03/2024
1° wedge window easy-
to-clean version "ETC"
Mount flange on the pro-
cess flange with another
graphite gasket between
flange and purging tube.
Screw on M12 nuts with
pads and tighten with 8
mm hex key and 19 mm
wrench. Tighten nuts with
torque (approximately 60
... 70 Nm) for good seal
on the gaskets.
5.2 Installing process flanges
Mounting gasket on the purging tube
Flange with window
ETC flange with window
M16 Screws mounting
Installing/mounting
19