Siemens SITRANS TDL Compact Operating Instructions page 42

In situ laser gas analyzers
Hide thumbs Also See for SITRANS TDL:
Table of Contents

Advertisement

Commissioning
7.4 Device startup
Table 7- 1
Startup Procedure Number
0
1
2
3
4
5
6
7
8
9
10
If the startup procedure stopped at the step below ten for more than five minutes, please
refer to the Startup Procedure troubleshooting steps.
Table 7- 2
Startup Procedure Number
0 to 6
7
1)
8 to 9
10
Contact Technical support if all other troubleshooting fails.
1)
42
Startup procedure steps
Startup procedure troubleshooting steps
1)
Information
Low level platform is not working
Device initialization and memory test
Internal communication check
Automatic gain control calibration procedure
Device parameters check and set
Lasers temperature stabilization (laser off)
Lasers temperature stabilization (laser on)
Lasers transmission check
Calibration check procedure
Concentration measurement check
Normal operation of the device (startup procedure is
finished)
Troubleshooting
Internal error occurred.
Please contact Siemens technical support.
No/Low optical transmission.
Please check the cleanliness of the process win-
dows and/or the flange alignment. As well, check
that the purging tubes are free.
Self-calibration procedure error. Please check the
device specification (device is used out of the
specification and/or the process gas is required for
reference purposes).
Normal operation of the device.
Compact Operating Instructions, 05\2023, A5E51659183-AA – revised 03/2024
SITRANS TDL