Operation
parent (range) screen. Click on "Back arrow" on this screen, will move
to range screen without saving edited values.
High: Range high edit box, click on edit box to enter high Range
PPM values.
RANGE HIGH edit screen:
Suppressed range: push button to toggle ACTIVE/DEACTIVE
(type invert/non invert).
The Save
button will save newly entered values.
4.7 Alarm Function
The Model 2000RS is equipped with 2 fully adjustable
concentration alarms and a system failure alarm. Each alarm has a relay
with a set of form "C" contacts rated for 1amp resistive load at 250
VAC. See Figure in Chapter 3, Installation and/or the Interconnection
Diagram included at the back of this manual for relay connections.
The system failure alarm has a fixed configuration described in
Chapter 3 Installation.
Teledyne Analytical Instruments
2000RS
80